Review the Reaction Setup

In the Reaction Setup screen, select the assay type (if using TaqMan reagents), then
review the calculated volumes for preparing the PCR reactions and sample dilutions. If
needed, you can edit the reaction volume, excess reaction volume, component
concentrations, and/or diluted sample concentration.
About the
In the comparative C
Example
• Applied Biosystems TaqMan
Experiment
• The reaction volume per well is 50 µL.
• The excess reaction volume is 10%.
• The reaction components are:
• The diluted sample concentration is 50 ng/µL.
• The sample stock concentration is 100 ng/µL.
Complete the
IMPORTANT!
Reaction Setup
Screen
Complete the Reaction Mix Calculations Tab for the Assays
1. Select the Reaction Mix Calculations tab (default).
2. In the Select Target pane, select 18S.
3. In the Assay Type drop-down list, select Inventoried/Made to Order.
4. Make sure the Reaction Volume Per Well field displays 50 µL.
5. Make sure the Excess Reaction Volume field displays 10%.
6. In the Reactions for 18S pane:
Notes
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
Experiments
T
example experiment:
T
®
– TaqMan
Universal PCR Master Mix (2✕)
– Assay Mix (20✕) for each target (18S, EGR3, MAOB, OGDH, OSGEP, and
SERPING1)
– Sample
– Water
Perform the following steps for each target assay in the reaction plate.
Make sure the Master Mix Concentration field displays 2.0✕.
a.
Make sure the Assay Mix Concentration field displays 20.0✕.
b.
Chapter 6 Design the Comparative C

Review the Reaction Setup

®
Gene Expression Assays are used.
Experiment
T
127