Planmeca ProMax 3D Plus User Manual page 62

With protouch x-ray unit
Hide thumbs Also See for ProMax 3D Plus:
Table of Contents

Advertisement

SETTINGS
58 Planmeca ProMax 3D Plus & 3D Mid with ProTouch
To control audio settings:
1. Select User > 1300 Operational Settings > 1320 Audio
Settings.
2. Use the minus or plus button to reduce or increase the
volume of the radiation warning tone.
3. Select the green check mark button.
To manage settings for patient positioning:
1. Select User > 1300 Operational Settings > 1330
Patient Positioning.
2. Turn an option(s) ON or OFF:
Side Entry:
Turn this option OFF if you do not wish to use the
open (full view) patient entry position. This might be
necessary if there is no space for the C-arm to
move back.
NOTE:
Planmeca ProMax 2D S2 units offer only one
patient entry position. This entry position is the
closed patient entry position where the C-arm is
positioned around the temple supports.
Midsagittal and Frankfort Lights in Tomo:
Turn this option ON if you wish to use the
midsagittal and Frankfort plane lights when you
take 3D exposures.
Temple Supports:
Turn this option OFF if you do not wish to use the
temple supports.
Return C-arm Automatically:
Turn this option ON if you wish to set the X-ray unit
so that the C-arm will automatically return to the
starting position at the end of an exposure cycle.
Note, however, that the automatic function works
only if the exposure button is pressed and held
down for the entire duration of the exposure.
User's Manual (3D)

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ProMax 3D Plus and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Promax 3d mid

Table of Contents