Cleaning; Service - Planmeca ProMax 3D Plus User Manual

With protouch x-ray unit
Hide thumbs Also See for ProMax 3D Plus:
Table of Contents

Advertisement

15

CLEANING

3D head supports
(incl. support bars and
fastening straps)
Chin cup / rest /
support
Bite pieces
Temple supports
Cephalostat head
supports
Patient handles
Other surfaces
(incl. touch screen)
ProFace sensor;
laser windows in the middle
ProFace sensor;
other surfaces (incl. glass
windows on both sides)
Other sensors
16

SERVICE

User's Manual (3D)
NOTE Disconnect the X-ray unit from mains before cleaning.
NOTE Do not use cleaning solutions in aerosol or spray form
directly on unit surfaces.
X-ray unit
soft cloth using
Autoclave
alcohol based
up to 135°C
cleaning solution
X
X
X
X
Sensors
Wipe with soft cloth
(NO CLEANING
SOLUTION)
X
X
To guarantee patient and user safety and to ensure
consistent image quality the X-ray unit must be checked
and recalibrated by a qualified Planmeca service
technician once a year or after every 10 000 exposures if
this is sooner.
Planmeca ProMax 3D Plus & 3D Mid with ProTouch 69
Wipe with
Wipe with
soft cloth using
mild cleaning
X
X
X
X
X
X
Compressed air
CLEANING
solution
X
X
X
X
X
X
X
X