To perform tube head seasoning:
1. Select User > 1500 Testing Routines > 1520 Tube
Head Seasoning.
This option allows you to warm up the X-ray tube, i.e.
run a tube head seasoning process. This is necessary
if the X-ray unit has not been used for a few days and
you receive error message E332 (Arcing across X-ray
tube).
2. Select the Start button.
3. Move to a protected area.
4. Press the exposure button when the word Ready
appears.
You can press and hold down the exposure button for
the whole duration of the process or remove your
finger from the exposure button when the word Wait
appears. Note that the seasoning process will take
several minutes. After a successful process the
message OK is shown.
5. Select the green check mark button.
NOTE Contact your service technician for help if error
message E332 (Arcing across X-ray tube) reappears
after a successful seasoning process.