Verifying Pulse Capture - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

8-44. Verifying Pulse Capture

Output @
SCOPE
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press
Each option in the PULSE menu is described below:
OUTPUT @ SCOPE Indicates the location of the signal output. If the signal does
not appear on the oscilloscope, press
AMPL Indicates the output level. You can select 2.5 V, 1.0 V, 250 mV, 100 mV, 25
mV, or 10 mV.
TRIG If you are using the external trigger, use this key to cycle through the trigger
settings. The available trigger settings are: off, /1 (trigger signal appears on each
marker), /10 (trigger signal appears on every tenth marker), and /100 (trigger signal
appears at every 100th marker).
You can also toggle the trigger off and on by pressing
MODE Indicates you are in PULSE mode. Use the softkey to change modes and
open menus for other oscilloscope calibration modes.
Default Pulse settings are 100.0 ns width and 1.000 ms period. To change these values,
you have several options. Usually, you will enter values for both pulse width and period.
Do this by entering the pulse width value with units first, followed immediately by the
period value and units, followed by
width of 50 ns and a period of 200 ns with the following sequence:
To change only the pulse width, enter a value in seconds. You can enter this value with
units (e.g., 200 ns) or without units (e.g., 0.0000002). To change only the period, enter a
frequency with units (e.g., 20 MHz, changing the period to 50 ns).
AMPL
2.5V
2.5V
1.0V
250 mV
100 mV
25 mV
10 mV
to return directly to the OTHER modes menu.
. . For example, you could enter a pulse
TRIG
MODE
off
pulse
off
volt
/1
edge
/10
levsine
/100
marker
wavegen
video
pulse
meas Z
overld
. To disconnect the signal, press
.
5500A-SC600 Option
Verifying Pulse Capture
.
8
8-31

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents