Shut Down Mode; Cjc; Celcius And Fahrenheit (°C And °F); Frequency Pulse Output Voltage - Fluke 726/FPC User Manual

Multifunction process calibrator
Table of Contents

Advertisement

Shut Down Mode

The Calibrator comes with a shut down mode set for 30
minutes (displayed for about 1 second when the
Calibrator is initially turned on). When shut down mode is
enabled, the Calibrator automatically shuts down after
the elapsed time from when the last key was pressed.
1.
Press C until SHUT DOWN appears on the
display.
2.
Use X and W to increase or decrease the time.
3.
Press S to save the setting.

CJC

Cold Junction Compensation (CJC) is a value for the cold
end of a thermocouple at the Meter's end.
1.
Press C until SELECT CJC appears on the
display.
2.
Use
and
Y
3.
Press S to save the setting.
Celcius and Fahrenheit ( ° C and ° F)
1.
Press C until SELECT UNIT °C (or °F) appears
on the display.
2.
Use X and W to select °C or °F.
3.
Press S to save the setting.
information@ITM.com
to select ON or OFF.
Z
Fluke-Direct
Multifunction Process Calibrator

Frequency Pulse Output Voltage

1.
Press C until FREQ OUTPUT V Adjust appears
on the display.
2.
Use X, W,
and
to adjust the frequency pulse
Y
Z
output voltage from 1 to 20 V.
3.
Press S to save the setting.

Pulse Output Frequency

1.
Press C until PULSE OUTPUT Hz FREQ Adjust
appears on the display.
2.
Use X, W,
and
to adjust the pulse output
Y
Z
frequency from 2 CPM to 15 kHz.
3.
Press S to save the setting.
.ca
Configuration Menus
15
1.800.561.8187

Advertisement

Table of Contents
loading

This manual is also suitable for:

726

Table of Contents