8-27. The V/DIV Menu

The V/DIV menu, shown below, sets the number of volts denoted by each division on
the oscilloscope. This menu provides alternative methods for changing the output
amplitude that may be more convenient for certain oscilloscope applications. To access
the V/DIV menu, press V/DIV from the VOLT menu.
20.00 mV/div
up
|
----------------
1 mV
2 mV
5 mV
10 mV
20 mV
50 mV
100 mV
200 mV
Each item in the V/DIV menu is described below:
V/div Changes the number of volts per division in the Output Display so that the
values selected correspond to the oscilloscope's input sensitivity (VOLTS/DIV.) The
available settings, shown in the figure above, are provided in 1-2-5 step increments.
Press the softkey under UP to increase the volts per division. Press the softkey under
DOWN to decrease the volts per division.
#DIV Specifies the number of divisions that establish the peak-to-peak value of the
waveform. The value can be adjusted from one to eight divisions. The amount
denoted by each division is displayed in the V/div field. Press the softkey under UP
to increase the signal's height, and press the softkey under DOWN to decrease it.

8-28. Shortcuts for Setting the Voltage Amplitude

The
and
oscilloscope in a 1-2-5 step sequence. For example, if the voltage is 40 mV, pressing
increases the voltage to the nearest cardinal point, which is 50 mV. Pressing
decreases the voltage to the nearest cardinal point, which is 20 mV.
Calibrating the Voltage Amplitude on an Oscilloscope
#DIV
down
up
|
|
----------------
|
0.5V
1V
2V
5V
10V
20V
50V
100V
keys step the voltages through cardinal point values of an
=
1
MODE
down
volt
|
volt
edge
|
levsine
1
marker
2
wavegen
3
video
4
pulse
5
meas Z
6
overld
7
8
5500A-SC600 Option
8
8-17