Oscilloscope Connections; Starting The Sc600 Option - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

8-17. Oscilloscope Connections

Using the cable supplied with the SC600 Option, connect the SCOPE output on the
Calibrator to one of the channel connectors on your oscilloscope (see Figure 8-1).
To use the external trigger, connect the TRIG OUT output on the Calibrator to the
external trigger connection on your oscilloscope. To use the external trigger and view its
signal with the calibration signal, connect the TRIG OUT output to another channel. See
your oscilloscope manual for details on connecting and viewing an external trigger.
Figure 8-1. Oscilloscope Connection: Channel and External Trigger

8-18. Starting the SC600 Option

Press
(LED lit) to select the SC600 Option. The SCOPE menu, shown below,
appears in the Control Display. You can press any of the first four softkeys to go directly
to the VOLT, EDGE, LEVSINE, and MARKER calibration menus. Press the last softkey
to go to the OTHER menu (also shown below), allowing access to WAVEGEN, VIDEO,
PULSE, Impedance/Capacitance measurement (MEAS Z), and Overload (OVERLD)
menus. Press
describes each of these menus in detail.
VOLT
WAVEGEN
to return to the SCOPE menu from the OTHER menu. This chapter
Select SCOPE mode:
EDGE
LEVSINE
------------
|
Select SCOPE mode (or PREV MENU):
VIDEO
PULSE
Oscilloscope Connections
NORMAL
V,
,
RTD
1000V
RMS
MAX
20V PK
MAX
TC
MARKER
OTHER
|
---------------
------
MEAS Z
OVERLD
5500A-SC600 Option
5500A
CALIBRATOR
SCOPE
AUX
200V PK
A,
-SENSE,
MAX
AUX V
HI
20V
RMS
MAX
TRIG
LO
OUT
1V PK
MAX
20V PK
MAX
gl001f.eps
8
8-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents