3. Set your oscilloscope's time base to show 10 time markers. The time markers should
4. Repeat this procedure for all time marker values recommended for your
5. Remove the signal by pressing

8-42. Testing the Trigger

The oscilloscope's ability to trigger on different waveforms can be tested using the wave
generator. When the wave generator is used, a square, sine, or triangle wave is
transmitted and the wave's output impedance, offset, and voltage can be varied in order
to test the triggering capability at different levels.
The wave generator should not be used for checking the accuracy of your
oscilloscope.
8-28
You may enter the equivalent frequency instead of the time marker value.
For example, instead of entering 200 ns, you may enter 5 MHz.
align with the oscilloscope divisions, as shown in the example below.
For an accurate reading, align the signal's peaks with the horizontal center axis.
oscilloscope. Repeat for digital and analog mode as required. Some oscilloscopes
may need the magnification changed while calibrating in analog mode.
Note
.
Note
Peaks are aligned
with center axis
gl011i.eps