Fluke 5500A Operator's Manual page 140

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual
4-56
c. Set the calibrator to 35.0 mA and press
d. Use the output adjustment controls to adjust the calibrator output for a reading of
+35.00 mA on the DMM. Verify that the error shown on the control display is
within specification.
e. Repeat using 350.0 mA. Verify the error is within specification.
f. Press the blue button on the DMM to switch to ac current measurement.
g. Set the calibrator output to 35.0 mA at 60 Hz. Verify the error is within
specification.
h. Repeat the previous step with the following calibrator settings:
AC Current
35.0 mA
350.0 mA
350.0 mA
i.
Press
on the calibrator and switch the DMM function switch to
j.
Set the calibrator output to 350 A at 0 Hz. and press
within specification.
k. Repeat the previous step using 3500 A at 0 Hz.
l.
Press
on the calibrator and press the blue button on the DMM to switch to
ac measurements.
m. Set the calibrator output to 350.0 A at 60 Hz and press
within specification.
n. Repeat the previous step with the following calibrator settings:
AC Current
350.0 A
3500.0 A
3500.0 A
.
Frequency
1.0 kHz
60 Hz
1.0 kHz
. Verify the error is
Frequency
1.0 kHz
60 Hz
1.0 kHz
.
. Verify the error is