Fluke 5500A Operator's Manual page 35

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

AC Voltage (Sine Wave) Specifications (cont.)
Range
Frequency
45 Hz to 1 kHz
100 to 1020 V
1 to 20 kHz
20 to 30 kHz
100 to 750 V
30 to 100 kHz
10 to 20 Hz
20 to 45 Hz
10 to 329.999 mV
45 Hz to 1 kHz
1 to 5 kHz
5 to 10 kHz
10 to 20 Hz
20 to 45 Hz
0.33 to 3.29999 V
45 Hz to 1 kHz
1 to 5 kHz
5 to 10 kHz
[1]
Remote sensing is not provided. Output resistance is < 5 m for outputs
maximum load capacitance is 500 pF, subject to the maximum burden current limits.
[2]
There are two channels of voltage output. The maximum frequency of the dual output is 10 kHz.
Absolute Uncertainty, tcal ± 5 C
± (% of output + V)
90 days
5725A Amplifier
0.04
80 mV
0.06
100 mV
0.08
100 mV
0.38
500 mV
Auxiliary Output [dual output mode only]
0.15
370
0.08
370
0.08
370
0.15
450
0.3
450
0.15
450
0.08
450
0.1
0.07
450
0.09
0.15
1400
0.2
0.3
1400
0.4
Introduction and Specifications
Resolution
1 year
0.05
80 mV
0.08
100 mV
10 mV
0.10
100 mV
0.5
500 mV
[2]
0.2
370
0.1
370
1 V
0.1
370
0.2
450
0.4
450
0.2
450
450
10 V
450
1400
1400
0.33 V. The AUX output resistance is < 1
General Specifications
Maximum
[1]
Burden
50 mA
70 mA
70 mA
70 mA
5 mA
5 mA
. The
1-15
1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents