Dc Power Amplitude Accuracy (Normal); Dc Power Amplitude Accuracy (Aux) - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual

7-19. DC Power Amplitude Accuracy (NORMAL)

This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The DC Power Amplitude Accuracy (NORMAL) tests the amplitude accuracy of the dc
power of the NORMAL terminals.
Nominal Value
(V)
(NORMAL)
20 mV
20 mV

7-20. DC Power Amplitude Accuracy (AUX)

This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The DC Power Amplitude Accuracy (AUX) tests the amplitude accuracy of the dc power
output of the AUX terminals.
Nominal Value
(V)
(NORMAL)
1000 V
1000 V
329 V
1000 V
7-18
Nominal Value
Measured Value
(A)
(AUX)
(NORMAL)
2.19 A
11 A
Nominal Value
Measured Value (A)
(A)
(AUX)
100 A
1 mA
2.19 A
11 A
Note
Deviation %
(V)
Note
Deviation %
(AUX)
90-Day Spec. (%)
0.020%
0.020%
90-Day Spec. (%)
0.06%
0.015%
0.025%
0.041%