1-14. Specifications

The following paragraphs detail specifications for the 5500A Calibrator. The specifications are valid after allowing a
warm-up period of 30 minutes, or twice the time the 5500A has been turned off. For example, if the 5500A has been
turned off for 5 minutes, the warm-up period is 10 minutes.
All specifications apply for the temperature and time period indicated. For temperatures outside of tcal + 5 C (tcal is the
ambient temperature when the 5500A was calibrated), the temperature coefficient is less than 0.1 times the 90-day
specifications per C (limited to 0 C to 50 C). These specifications also assume the 5500A Calibrator is zeroed every
seven days or when the ambient temperature changes more than 5 C. (See "Zeroing the Calibrator" in Chapter 4 of the
5500A Operator Manual.)
Also see additional specifications later in this chapter for information on extended specifications for ac voltage and
current. The dimensional outline for the 5500A Calibrator is shown in Figure A.
5500A
CALIBRATOR
SCOPE
NORMAL
AUX
V,
,
A,
-SENSE,
200V PK
RTD
AUX V
STBY
MAX
HI
7
1000V
20V
RMS
RMS
MAX
MAX
TRIG
LO
OUT
4
1V PK
MAX
20V PK
MAX
1
TC
/ +
20V PK
MAX
43.2 cm (17 in)
PREV
OPR
EARTH
SCOPE
BOOST
MENU
µ
dBm
sec
8
9
SETUP
RESET
m
V
Hz
n
W
¡F
NEW
5
6
CE
k
A
¡C
REF
p
MEAS
TRIG
2
3
M
F
TC
OUT
MULT
DIV
0
SHIFT
ENTER
x
47.0 cm (18.5 in)
Figure A. 5500A Calibrator Dimensional Outline
Introduction and Specifications
17.8 cm
(7 in)
EDIT
FIELD
POWER
I
O
Specifications
6.4 cm
(2.5 in)
For Cable
Access
om002f.ewps
1
1-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents