Fluke 5500A Operator's Manual page 227

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

x
*PUD?
IEEE-488
Returns the contents of the *PUD (Protected User Data) memory in definite length
format.
Response:
#2nn<nn characters>
Example:
*PUD? returns #216CAL LAB NUMBER 1
Return #2 then 16 then 16 characters of text (including spaces) stored in the nonvolatile
memory.
x
RANGE?
IEEE-488
(Range query) Return the present output ranges. Both the primary output and secondary
outputs are returned. If there is no secondary output, 0 is returned. Dual outputs are
noted with P for primary output (front panel NORMAL terminals) and S for secondary
output (front panel AUX terminals).
Response:
<primary output>,<secondary output>
Examples:
DC330MV,0
DC33MA_A,0
AC3_3V,0
AC330MA_A,0
R110OHM,0
C1_1UF,0
TCSRC,0
RTD_330,0
DC3_3V_P,DC2_2A_AS
AC330V_P,AC11A_AS
DC330MV_P,DC3_3V_S
AC330V_P,AC3_3V_S
Return the symbolic name of the single or first output, and return the symbolic name of
the second output (0 if there is no second output).
x
RANGELCK
IEEE-488
(Range Lock command) Locks in the present range, or selects auto ranging for dc
voltage and dc current single outputs. The range automatically unlocks if the output
function changes, for example from dc volts to dc current. When RANGELCK is on, this
is equivalent to the softkey range lock showing locked. When RANGELCK is off, this is
equivalent to the softkey range lock showing auto.
Parameter:
ON
OFF
Example:
RANGELCK OFF
Set the range lock off to allow autoranging for dc volts or dc current.
x
x
RS-232
Sequential
x
x
RS-232
Sequential
(dc volts 330 mV range)
(dc current 33 mA range)
(ac volts 3.3 V range)
(ac current 330 mA range)
(ohms 110
(capacitance 1.1 F range)
(temperature thermocouple source)
(temperature RTD 330
(dc power 3.3 V, 2.2 A ranges)
(ac power 330 V, 11 A ranges)
(dual dc volts 330 mV, 3.3 V ranges)
(dual ac volts 330 V, 3.3 V ranges)
x
x
RS-232
Sequential
(Locks the dc volts or dc current range)
(Unlocks the dc volts or dc current range for autoranging)
Remote Commands
Summary of Commands and Queries
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
range)
range)
x
x
Overlapped
6
Coupled
6-31

Hide quick links: