Dc Current Amplitude Accuracy - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

7-9.

DC Current Amplitude Accuracy

The DC Voltage Amplitude Accuracy test verifies the accuracy of dc current at the
5500A Calibrator front panel AUX terminals.
Range
Nominal
Value
3.3 mA
0 mA
3.3 mA
0.19 mA
3.3 mA
-0.19 mA
3.3 mA
1.9 mA
3.3 mA
-1.9 mA
3.3 mA
3.29 mA
3.3 mA
-3.29 mA
33 mA
0 mA
33 mA
19 mA
33 mA
-19 mA
33 mA
32.9 mA
33 mA
-32.9 mA
330 mA
0 mA
330 mA
190 mA
330 mA
-190 mA
330 mA
329 mA
330 mA
-329 mA
2.2 A
0A
2.2 A
2.19 A
2.2 A
-2.19 A
11 A
0 A
11 A
11 A
11 A
-11 A
Measured Value (A)
Performing a Calibration Check
90-Day Spec.
Deviation %
(% or mA)
0.00005 mA
0.036%
0.036%
0.013%
0.013%
0.012%
0.012%
0.00025 mA
0.009%
0.009%
0.009%
0.009%
0.0033 mA
0.010%
0.010%
0.009%
0.009%
0.000044A
0.025%
0.025%
0.00033A
0.041%
0.041%
Maintenance
7
7-9