Dc Voltage Amplitude Accuracy (Aux) - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual
7-8.

DC Voltage Amplitude Accuracy (AUX)

The DC Voltage Amplitude Accuracy test verifies the accuracy of dc voltage at the
5500A Calibrator front panel AUX terminals in the presence of a lower voltage at the
NORMAL terminals.
Nominal Value
(NORMAL)
3 V
3 V
3 V
3 V
3 V
3 V
7-8
Nominal Value
Measured Value
(AUX)
0 mV
329 mV
-329 mV
0.33 V
3.29 V
-3.29 V
Deviation %
(AUX V)
90-Day Spec.
(% or mV)
0.350 mV
0.1365%
0.1365%
0.1361%
0.0407%
0.0407%