Fluke 5500A Operator's Manual page 207

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

x
CAL_STATE?
IEEE-488
(Calibration State query) Return state of calibration.
Response:
RON - running a calibration step
REF - waiting for a CAL_NEXT with reference (measurement) value
INS - instruction available, waiting for a CAL_NEXT
NOT - not in a calibration procedure (or at end of one)
x
CAL_STEP?
IEEE-488
(Calibration Step query) Return name of step currently running.
Response:
(chart) The step name
Example:
IDAC_RATIO (running IDAC ratio calibration)
NOT (not running a calibration procedure now)
x
CAL_STORE
IEEE-488
(Calibration Store command) Store new calibration constants (CAL switch must be
ENABLED).
Parameter:
(optional) (int) Date to be stored with the constants
Other than requiring that it be a single integer between 2,147,483,648;
the 5500 attaches no significance to the date stored. It is printed in any
calibration reports and may be fetched using CAL_DATE?
Example:
CAL_STORE 081294 - could mean August 12, 1994
x
CAL_STORE?
IEEE-488
(Calibration Store query) return whether a cal store is needed.
Response:
1 is yes, 0 if no
x
CAL_SW?
IEEE-488
(Calibration Switch query) Return the setting of the calibration enable switch.
Response:
(integer) 1 for enable, 0 for normal
Example:
1
x
x
RS-232
Sequential
x
x
RS-232
Sequential
x
x
RS-232
Sequential
x
x
RS-232
Sequential
x
x
RS-232
Sequential
Remote Commands
Summary of Commands and Queries
x
x
Overlapped
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
x
x
Overlapped
x
x
Overlapped
Coupled
6
Coupled
Coupled
6-11

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents