1-28. Calculating Power Uncertainty

Overall uncertainty for power output in Watts (or VARs) is based on the root sum square (rss) of the individual
uncertainties in percent for the selected voltage, current, and power factor parameters:
Watts uncertainty
U
power
VARs uncertainty
U
U
VARs
Because there are an infinite number of combinations, you should calculate the actual ac power uncertainty for your
selected parameters. The method of calculation is best shown in the following examples (using 90-day specifications):
Example 1 Output: 100 V, 1 A, 60 Hz, Power Factor = 1.0 ( =0)
Voltage Uncertainty Uncertainty for 100 V at 60 Hz is 0.04 % + 6.6 mV, totaling:
100 V x .0004 = 40 mV added to 6.6 mV = 46.6 mV. Expressed in percent:
46.6 mV/100 V x 100 = 0.047 % (see "AC Voltage (Sine Wave) Specifications").
Current Uncertainty Uncertainty for 1 A is 0.08 % + 300 A, totaling:
1 A x .0008 = 800 A added to 300 A = 1.1 mA. Expressed in percent:
1.1 mA/1 A x 100 = 0.11 % (see "AC Current (Sine Waves) Specifications").
PF Adder Watts Adder for PF = 1 ( =0) at 60 Hz is 0 % (see "Phase Specifications").
Total Watts Output Uncertainty =
Example 2 Output: 100 V, 1 A, 400 Hz, Power Factor = 0.5 ( =60)
Voltage Uncertainty Uncertainty for 100 V at 400 Hz is 0.04 % + 6.6 mV, totaling:
100 V x .0004 = 40 mV added to 6.6 mV = 46.6 mV. Expressed in percent:
46.6 mV/100 V x 100 = 0.047 % (see "AC Voltage (Sine Wave) Specifications").
Current Uncertainty Uncertainty for 1 A is 0.08 % + 300 A, totaling:
1 A x .0008 = 800 A added to 300 A = 1.1 mA. Expressed in percent:
1.1 mA/1 A x 100 = 0.11 % (see "AC Current (Sine Wave) Specifications").
PF Adder Watts Adder for PF = 0.5 ( =60) at 400 Hz is 2.73 % (see "Phase Specifications").
Total Watts Output Uncertainty =
VARs When the Power Factor approaches 0.0, the Watts output uncertainty becomes unrealistic because the dominant
characteristic is the VARs (volts-amps-reactive) output. In these cases, calculate the Total VARs Output Uncertainty, as
shown in example 3:
Example 3 Output: 100 V, 1 A, 60 Hz, Power Factor = 0.0872 ( =85)
Voltage Uncertainty Uncertainty for 100 V at 60 Hz is 0.04 % + 6.6 mV, totaling:
100 V x .0004 = 40 mV added to 6.6 mV = 46.6 mV. Expressed in percent:
46.6 mV/100 V x 100 = 0.047 % (see "AC Voltage (Sine Wave) Specifications").
Current Uncertainty Uncertainty for 1 A is 0.08 % + 300 A, totaling:
1 A x .0008 = 800 A added to 300 A = 1.1 mA. Expressed in percent:
1.1 mA/1 A x 100 = 0.11 % (see "AC Current (Sine Wave) Specifications").
VARs Adder VARs Adder for
Total VARS Output Uncertainty =
2
2
2
U
U
U
voltage
current
PFadder
2
2
2
U
U
voltage
current
VARsadder
2
2
2
U
. 0
047
. 0
11
0
power
2
2
U
. 0
047
. 0
11
. 2
73
power
=85 at 60 Hz is 0.02 % (see "Phase Specifications").
2
2
U
=
. 0
047
. 0
11
. 0
VARs
Introduction and Specifications
. 0
12
%
2
. 2
73
%
2
02
. 0
12
%
General Specifications
1-25
1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents