Testing Harmonics Volts Performance; Harmonics Performance For Volts, Harmonics Screen - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual
4. Press the WAVE MENUS softkey on the calibrator. Ensure the Phase angle is 0.00
5. Select W from VAW on the Tester.
6. Press M until the Tester is in the Text Screen mode. Verify that the W/KW,
7. Press M until the Tester is in the Harmonics screen mode. Verify that the
8. Repeat the previous three steps using the calibrator outputs and performance limits
9. Press

4-60. Testing Harmonics Volts Performance

1. Press M until the harmonics screen is displayed on the Tester.
2. Press F until V is displayed above the upper right corner of the harmonics display.
3. Press f until AP is displayed in the top status line of the Tester.
4. Press S until %20s is displayed in the top status line of the Tester.
5. Connect the calibrator NORMAL output to the V and COM connectors on the
6. Connect the calibrator AUX output to the Current Probe connector on the Tester.
7. Set the calibrator output to 7.0V at 60 Hz on the NORMAL output and 700 mV at
8. Move the Tester cursor to the corresponding harmonic number.
9. Verify that the harmonic amplitude and phase angle readings displayed by the Tester
10. Repeat the previous three steps using the settings and limits in Table 4-5.
Amplitude
11. Press
4-60
degrees. Press
.
VA/KVA, and VAR/KVAR readings are within the minimum and maximum limits
specified in Table 4-4.
fundamental frequency phase angle readings are between the minimum and
maximum readings listed in Table 4-4.
listed in Table 4-4.
to remove the voltage from the Tester.
Tester.
60-Hz on the AUX output. Press the WAVE MENUS softkey and ensure the phase
angle is -10.0 degrees. Press the HARMONIC MENU softkey and ensure the
HARMONIC selection is set to "1" and the FUNDMTL selection is set to "aux."
Press
.
are within the minimum and maximum limits listed in Table 4-5. (Note: The Tester
will read a positive phase when the 5500A output is a negative phase because, on the
5500A, the polarity of the phase is always relative to the NORMAL channel output.)
Table 4-5. Harmonics Performance for Volts, Harmonics Screen
5500A
Normal Output
Harmonic
(V)
No.
7.00
1
7.00
3
7.00
9
7.00
13
7.00
21
7.00
31
to remove the voltage from the Tester.
Fluke
Tester
Harmonic
Phase
cursor
(deg.)
No.
MIN
-10
1
-20
3
-30
9
-40
13
-50
21
-60
31
Performance Limits
Amplitude
Phase
MAX
MIN
6.7
7.3
8
6.7
7.3
14
6.7
7.3
21
6.7
7.3
29
6.5
7.5
35
6.2
7.8
40
MAX
12
26
39
51
65
80