5560A/5550A/5540A
Calibrator
Service Manual
September 2023
© 2023 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 5560A

  • Page 1 5560A/5550A/5540A Calibrator Service Manual September 2023 © 2023 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 2 Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
  • Page 3: Table Of Contents

    Volts DC Calibration (OUTPUT VZ)............12 Volts DC Calibration (AUX Output) (5550A and 5560A)....... 15 Amps DC Calibration ................17 Resistance Calibration................26 Volts AC Calibration (OUTPUT VZ)............33 Volts AC Calibration (AUX Output) – 5550A and 5560A ...... 39 Amps AC Calibration ................41...
  • Page 4 Service Manual Capacitance..................53 Inductance (5550A and 5560A) ............57 Thermocouple Calibration..............58 Phase Calibration (5560A and 5550A) ..........61 Frequency Calibration................64 Calibration Adjustment ................65 Start the Calibration Adjustment ............66 Volts DC Calibration Adjustment (OUTPUT VZ)........66 Volts AC Calibration Adjustment (OUTPUT VZ)........
  • Page 5: Introduction

    Introduction The 5560A/5550A/5540A Calibrators (the Product or the Calibrator) addresses a wide calibration workload that includes 6.5 digit bench Digital Multimeters (DMMs), and comes with internal and external features that protect it against damage and make it easier to transport for on-site or mobile ®...
  • Page 6: Contact Fluke Calibration

    Fluke Corporation operates worldwide. For local contact information, go to our website: www.flukecal.com To register your product, view, print, or download the latest manual or manual supplement, go to our website. +1-425-446-5500 info@flukecal.com Product-Use Information For Product-use information, please see the 5540A/5550A/5560A Operators Manual available online at www.flukecal.com.
  • Page 7: Safety Information

    (IEEE-488.1) standard and supplemental IEEE-488.2 standard. Under the remote control of an instrument controller, the Product operates exclusively as a talker/listener. Use the IEEE-488 command set or run MET/CAL software (optional) to write your own programs. See the 5560A/5550A/ 5540A Remote Programmers Manual at www.flukecal.com...
  • Page 8: Remote Operation (Usbtmc)

    Calibrator and complies with supplemental IEEE-488.2 standard. The Ethernet port connects a host PC to the Product. To send commands to the Product, enter commands from a telnet session running on the host computer. See the 5560A/5550A/5540A Remote Programmers Manual at www.flukecal.com for a discussion of the Ethernet commands available for Ethernet operation.
  • Page 9: In-Guard

    Calibrator Theory of Operation In-guard From front to back, the in-guard assemblies are: • A15 terminal indicators • A14 output switching • A10 TC, A3 analog motherboard • Optional A4 scope option • A5 impedance • A6 DDS • A8 HV amplifiers •...
  • Page 10: A7 Current

    Performance limits specified in the test record tables in this section are based on the 90-day specifications for the 5560A at a 99 % level of confidence. For the 5550A, 5540A and/or if limits to other specifications are necessary, the test records must be modified. A description of how to determine a guardband test limit is included in this section.
  • Page 11: Warmup Procedure For All Verification Tests

    Calibrator Calibration Verification of the Product adequate to guarantee the performance of the Product for its functions and ranges. Warmup Procedure for All Verification Tests Notes Consider these grounding and guarding suggestions: • For capacitance and inductance, disconnect the Guard from the Ground strap. •...
  • Page 12 Table 1. Required Equipment for Main Output Equipment Description Manufacture/Model/PN Application AC Measurement Standard Fluke Calibration 5790B Vac, Iac Fluke 742A Series (100, 1 k, 10 k, Resistance Standards Ohms, Idc 100 k, 1 M, 10 M) DC Reference Standard Fluke 732C Frequency Counter...
  • Page 13: Determine Specification Limits For Other Calibration Intervals

    The verification procedures in this document tests to the 90-day, 99 % confidence specification limits of the 5560A. For the 5550A and 5540A or for other calibration intervals, you must calculate limits based upon the specification that was selected. The subsequent examples show how the 1-year specification limits are calculated.
  • Page 14: Applying Guardbands To Specification Limits

    These must be accounted for their individual situations, so it is not possible for Fluke Calibration to estimate uncertainty for all user calibrations. Those doing the calibration of the Product can verify to 99 % confidence limits, 95 % confidence limits, 90-day, 1-year specifications, or 2-year specifications.
  • Page 15 Calibrator Calibration Verification of the Product Figure 2. Test Limits Established by Guardbanding Meas + TL + SL There are many ways to develop test limits from decision rules. Two of these rules are shown as examples to develop test limits through guardbanding. ILAC G8 (ISO 14253-1) Decision Rule In this guardband strategy example, determine the test limit by the subtracting the uncertainty of the measurement from the specification.
  • Page 16: Volts Dc Calibration (Output Vz)

    5730A-7003 To calibrate the Volts dc function from the main output: 1. On the Fluke Calibration 8588A put a 4-wire short (Fluke PN 2540973) across the HI and LO input and sense terminals. 2. Set the 8588A as follows: Volts dc, Range Auto, NPLC 50, Guard ON (external guard) 3.
  • Page 17 Figure 3. Volts DC Calibration Direct 8588A 5560A 22. For steps other than ±100 mV, ±1 V, and ±10 V the output of the Product is the same as the 8588A measurement recorded directly when connected to the product output DMM Reading on Product column.
  • Page 18 5560A/5550A/5540A Service Manual 24. With the entire table populated, evaluate each step for compliance to limits. If Guardbanding is to be applied create a table from the example in this manual, where the acceptance limits are properly modified by your Guardbanding method.
  • Page 19: Volts Dc Calibration (Aux Output) (5550A And 5560A)

    5730A-7003 To calibrate the Volts dc function from the secondary output: 1. On the Fluke Calibration 8588A put a 4-wire short (Fluke PN 2540973) across the HI and LO input and sense terminals. 2. Set the 8588A as follows: Volts dc, Range Auto, NPLC 50, Guard ON (external guard).
  • Page 20 5560A/5550A/5540A Service Manual Figure 4. Volts DC Calibration SENSE Output 5560A 8588A 7. Connect a strap between the Guard and Ground terminals of the DUT. 8. Output from the DUT/Measure with the meter and record the 8588A measured values when connected to the Product in the Product Output column, see Table 5.
  • Page 21: Amps Dc Calibration

    Table 7. 6. Record the shunt resistance in the Ω column marked as Shunt_R in Table 7. 7. Connect the 8588A for V dc measurement with the low thermal cable to Fluke 742A-1k SENSE See Figure 5.
  • Page 22 5560A/5550A/5540A Service Manual Figure 5. Amps DC with 742A-1k - Meter Zero 8588A 742A-1k 742A-1k Resistance Standard CURRENT SENSE NOMINAL VALUE = 1 K: TEMPERATURE RANGE: 18 °C TO 28 °C ‘ MAXIMUM VOLTAGE = 10 VDC MAXIMUM CURRENT = 0.01 A...
  • Page 23 15. Toggle the power switch of the A40B-1mA shunt to ON. Continue only if power light illuminates and the BATT LOW is off. 16. With no current applied from the Product to the Fluke A40B-1mA shunt, proceed with the zero meter function. On the 8588A, push ZERO, and then ZERO FUNC. Allow the zero function to finish.
  • Page 24 22. Set the Product to Standby. Disconnect the shunt. Connect the test equipment as shown in Figure 9. 23. With no current applied from the Product to the Fluke A40B-100mA shunt, proceed with the zero meter function. On the 8588A, push ZERO, and then ZERO FUNC. Allow the zero function to finish.
  • Page 25 25. Set the Product to Standby. Disconnect the shunt. Connect the test equipment as shown in Figure 10. 26. With no current applied from the Product to the Fluke A40B-1A shunt, proceed with the zero meter function. On the 8588A, push ZERO, and then ZERO FUNC. Allow the zero function to finish.
  • Page 26 35. Set the product to Standby. Disconnect the shunt. Connect the test equipment as shown in Figure 12. 36. With no current applied from the Product to the Fluke A40B-10A shunt, proceed with the meter zero function. On the 8588A, push ZERO, and then ZERO FUNC. Allow th zero function to finish.
  • Page 27 38. Set the Product to Standby. Disconnect the shunt. Connect the test equipment as shown in Figure 13. 39. With no current applied from the Product to the Fluke A40B-20A shunt, proceed with the meter zero function. On the 8588A, push ZERO, and then ZERO FUNC. Allow the zero function to finish.
  • Page 28 120 µA -10 µA Fluke 742A-1k ±7.0 nA 120 µA 100 µA Fluke 742A-1k ±16.0 nA 120 µA -100 µA Fluke 742A-1k ±16.0 nA 1.2 mA 0 mA Fluke 742A-1k ±15.0 nA 1.2 mA -0.000001 mA Fluke 742A-1k ±15.0 nA...
  • Page 29 Step Shunt on Shunt Product Limits Range Output [Ω] Output Output 1.2 mA 0.1 mA Fluke 742A-1k ±23.0 nA 1.2 mA -0.1 mA Fluke 742A-1k ±23.0 nA 1.2 mA 1 mA Fluke A40B-1mA ±95.0 nA 1.2 mA -1 mA Fluke A40B-1mA ±95.0 nA...
  • Page 30: Resistance Calibration

    (whenever the TUR is low). • Different resistance modes. Because of product compliance (Operating Characteristics) certain steps require the Fluke Calibration 8588A to be used in Low Current Mode (LoI) to reduce the current into the product. For all resistance steps: Toggle the meter to the Volts dc function prior to putting the Product in operate mode.
  • Page 31 Calibration Verification of the Product Table lists the required equipment. Table 8. Required Equipment for Resistance Equipment Description Manufacture/Model/PN Quantity Digital Multimeter Fluke Calibration 8588A Resistance Standard Fluke 742A-100k Resistance Standard Fluke 742A-1M Resistance Standard Fluke 742A-10M Resistance Standard MI 9331G/100 M...
  • Page 32 8588A 6. Set the Product to Standby. 7. Connect the 8588A to Fluke 742A-100k for 4-Wire measurement. 8. Set the 8588A as follows: Resistance 4Wire, LoI OFF, Range (as per Table 10), NPLC 50, Guard OFF (external guard light ring off).
  • Page 33 8588A measured value in the DMM Measurement on Product column in Table 10. 20. Set the product to Standby. 21. Connect the 8588A to Fluke 742A-1M for 2-Wire measurement. 22. Measure with the meter and record the 8588A measured value in the DMM Measurement on Transfer Resistor column in Table 10.
  • Page 34 5560A/5550A/5540A Service Manual Calculate the Meter error (Linear Interpolation) with this formula: -X)+Y (X-X where X =0.1 and X =1 and X is the value from the Product Output column in Table for each step. 34. Connect the existing setup used for high resistance to the product for 2-W measurements and complete steps 23 to 26 from Table 10.
  • Page 35 Calibrator Calibration Verification of the Product 38. Input the value of the 742A-10k resistor into the Transfer Resistor column steps 23 and 24 of in Ω . Table 39. Starting with step 23, output the resistance value from the Product. 40.
  • Page 36 5560A/5550A/5540A Service Manual Table 10. Calibration Steps for Resistance (cont.) Measure- Calculated Product Product DMM Mode/ Transfer Transfer Measure- Step ment on Product Limits Range Output Method Resistor Value ment on Transfer Output Product Resistor Ω Normal 1 k Ω...
  • Page 37: Volts Ac Calibration (Output Vz)

    Table lists the required equipment. Table 11. Required Equipment Volts AC Main Output Equipment Description Manufacture/Model/PN Quantity AC Measurement Standard Fluke 5790B Digital Multimeter Fluke 8588A Double banana RG58 cable Pomona Model 2BC-24 Single banana to single banana lead Various...
  • Page 38 5560A/5550A/5540A Service Manual The volts ac calibration uses two methods. For frequencies ≤ 10 Hz low frequency DMM (Fluke 8588A) is used. For frequencies >10 Hz the calibration is accomplished with Fluke 5790B AC Measurement Standard. 1. Set the 8588A as follows: Volts ac, Range Auto, Band Wideband, Coupling: DC 10M Ω , Guard ON (external guard).
  • Page 39 Calibrator Calibration Verification of the Product Figure 18. Volts AC Calibration 5560A 5590B 12. Starting with step 40, make a measurement with the 5790B. Record the measurement in the Product Output column of Table 12. 13. Continue with the rest of the steps for this connection (steps 40 to 121).
  • Page 40 5560A/5550A/5540A Service Manual Table 12. Calibration Steps for Volts AC (cont.) Calculated Step Product Range Product Output Product Frequency Product Limits Output 12 mV 10 mV 10 Hz ±7.2 µV 12 mV 10 mV 1 kHz ±7.2 µV 12 mV...
  • Page 41 Calibrator Calibration Verification of the Product Table 12. Calibration Steps for Volts AC (cont.) Calculated Step Product Range Product Output Product Frequency Product Limits Output 1.2 V 0.121 V 1 kHz ±74 µV 1.2 V 0.121 V 20 kHz ±74 µV 1.2 V 0.121 V 50 kHz...
  • Page 42 5560A/5550A/5540A Service Manual Table 12. Calibration Steps for Volts AC (cont.) Calculated Step Product Range Product Output Product Frequency Product Limits Output 12 V 10 V 5 Hz ±7.75 mV 12 V 10 V 10 Hz ±1.50 mV 12 V...
  • Page 43: Volts Ac Calibration (Aux Output) - 5550A And 5560A

    To proceed with volts ac out of the Sense VZ LO/Output VI AUX proceed as follows: 1. Connect the Product to Fluke 5790B as shown in Figure 19. 2. Connect a strap between the guard and ground terminals of the DUT.
  • Page 44 5560A/5550A/5540A Service Manual Figure 19. Volts AC Out of the SENSE VZ LO/OUTPUT VI AUX Calibration 5560A 5790B 3. Set the 5790B for Input 2, Auto range, High resolution ON, GUARD ON. 4. Start with step 1 in Table 14. Set the DUT to 1 V for the main output and Product output setting V and Hz in Table 14.
  • Page 45: Amps Ac Calibration

    The amps ac calibration uses two methods. For frequencies ≤ 10 Hz, use a low-frequency Fluke Calibration 8588A meter to monitor the output of Fluke A40B series shunts. For frequencies >10 Hz the calibration is accomplished with the Fluke Calibration 5790B AC Measurement Standard monitoring...
  • Page 46 5560A/5550A/5540A Service Manual To calibrate the amps ac function: 1. Set the 8588A as follows: Volts ac, Range Auto, Wide Bandwidth, Coupling: DC 10 M Ω , Guard ON (external guard). 2. Make sure that the DUT is in Standby.
  • Page 47 19. Connect the test equipment as shown in Figure 21. Figure 21. Amps AC with A40B-10mA 8588A 5560A A40B-10mA 20. From step 31, Table 16, output from the DUT/Measure with the meter and record the 8588A measurement in the DMM Reading on Shunt Output [V] column.
  • Page 48 5560A/5550A/5540A Service Manual Figure 22. Amps AC with A40B-100mA 8588A 5560A A40B-100mA OUTPUT 28. From step 43, Table 16, set the output from the DUT/Measure with the meter and record the 8588A measurement in the DMM Reading on Shunt Output [V] column.
  • Page 49 Calibrator Calibration Verification of the Product Figure 23. Amps AC with A40B-1A 8588A 5560A A40B-1A OUTPUT 36. From step 55, Table 16, set the output from the DUT and wait 30 seconds for the output to settle. Measure with the meter and record the 8588A measurement in the DMM Reading on Shunt Output [V] column.
  • Page 50 5560A/5550A/5540A Service Manual Figure 24. Amps AC with A40B-5A 8588A 5560A A40B-5A OUTPUT INPUT 50. Move the red lead connected to the Product VI AUX to the 30 A terminal. 51. Move the connection of the shunt output to the 8588A.
  • Page 51 64. Set the product to STBY. 65. Connect the test equipment as shown in Figure 26. Figure 26. Amps AC with A40B-20A 8588A 5560A A40B-20A OUTPUT INPUT 66. From step 81, Table 16, output from the DUT and wait 3 minutes for the output to settle. Measure with the meter and record the 8588A measurement in the DMM Reading on Shunt Output [V] column.
  • Page 52 5560A/5550A/5540A Service Manual Figure 27. Amps AC with A40B-50A A40B-50A 8588A 5560A OUTPUT INPUT 74. From step 89, Table 16, output from the DUT and wait 3 minutes for the output to settle. Measure with the meter and record the 8588A measurement in the DMM Reading on Shunt Output [V] column.
  • Page 53 Reading on Step Shunt Product Limits Range Output Frequency [Ω] Shunt Output Output [V] Fluke A40B- 120 μA 10 μA 3 Hz ±12.0 nA Fluke A40B- 120 μA 10 μA 45 Hz ±12.0 nA Fluke A40B- 120 μA 10 μA 1 kHz ±12.0 nA...
  • Page 54 5560A/5550A/5540A Service Manual Table 16. Calibration Steps for Amps AC (cont.) Calculated Product Product Product Shunt_R Reading on Step Shunt Product Limits Range Output Frequency [Ω] Shunt Output Output [V] Fluke A40B- 1.2 mA 0.121 mA 10 kHz ±245 nA Fluke A40B- 1.2 mA...
  • Page 55 Shunt_R Reading on Step Shunt Product Limits Range Output Frequency [Ω] Shunt Output Output [V] Fluke A40B- 12 mA 10 mA 1 kHz ±3.00 μA 10mA Fluke A40B- 12 mA 10 mA 5 kHz ±3.00 μA 10mA Fluke A40B- 12 mA...
  • Page 56 5560A/5550A/5540A Service Manual Table 16. Calibration Steps for Amps AC (cont.) Calculated Product Product Product Shunt_R Reading on Step Shunt Product Limits Range Output Frequency [Ω] Shunt Output Output [V] Fluke A40B- 1.2 A 0.121 A 3 Hz ±0.12 mA...
  • Page 57: Capacitance

    Product Limits Range Output Frequency [Ω] Shunt Output Output [V] 12 A 10 A 1 kHz Fluke A40B-10A ±2.90 mA 12 A 10 A 5 kHz Fluke A40B-10A ±3.80 mA 12 A 10 A 10 kHz Fluke A40B-10A ±21.0 mA 30.2 A...
  • Page 58 5560A/5550A/5540A Service Manual To calibrate the Capacitance function for values <1 mF: 1. Remove the Guard Ground strap on the Product. 2. Set the LCR meter for Cp Mode, Level 1V, Freq 1 kHz, Speed Slow. 3. Complete the OPEN/SHORT compensation of the LCR meter and lead set for two-wire measurement that includes the Y adapter and single banana plugs.
  • Page 59 A measurement procedure for capacitance is to apply a known current across the capacitor and then measure the voltage change for a known time interval. Fluke Calibration recommended to implement the following in computer-controlled routine for consistency of results and timely sequence execution.
  • Page 60 5560A/5550A/5540A Service Manual Figure 28. Capacitance Charge Connection 8588A 5730A 5560A The Calculated Product Output is calculated with this formula: Calculated Product Output = Test_current * 17. Use the appropriate conversion to convert from base units to units for the specific evaluation step.
  • Page 61: Inductance (5550A And 5560A)

    Calibrator Calibration Verification of the Product Inductance (5550A and 5560A) Table lists the required equipment. Table 20. Required Equipment for Inductance Calibration Equipment Description Manufacture/Model/PN Quantity LCR Meter Hioki IM3533 BNC (M) to BNC (M) 4W lead set for LCR...
  • Page 62: Thermocouple Calibration

    5560A/5550A/5540A Service Manual Table 21. Calibration Steps for Inductance with LCR Meter (cont.) Step Product Range Product Output Test Frequency Calculated Product Output Limits 120 mH 12.1 mH 100 Hz ±113 µH 120 mH 100 mH 100 Hz ±210 µH 1.2 H...
  • Page 63 2. Connect a copper thermocouple mini plug with a twisted pair copper wire to the Product thermocouple port. 3. Connect the twisted pair to the Fluke 5730A OUTPT HI/LO. Observe correct polarity. 4. Disconnect the V-Guard to Ground strap on the 5730A.
  • Page 64 1. Connect a strap between the Guard and Ground terminals of the DUT. 2. Connect a copper thermocouple mini plug with twisted pair copper wire to Fluke 8588A INPUT HI/ LO (observe proper polarity). Short the thermocouple mini plug under the VI AUX binding post of the 5560A.
  • Page 65: Phase Calibration (5560A And 5550A)

    12. With all steps in the table populated, evaluate each step for compliance to limits. If Guardbanding is to be applied, create a table from the example in this manual, where the acceptance limits are properly modified by your Guardbanding method. Phase Calibration (5560A and 5550A) Table lists the required equipment.
  • Page 66 5560A/5550A/5540A Service Manual To calibrate dual output V/V: 1. Connect the phase meter REFERENCE INPUT to the Product OUTPUT VZ HI/LO with the double banana RG58 cable and BNC (M) to the binding post adapter. 2. Connect the phase meter SIGNAL INPUT to the Product SENSE VZ LO/VI AUX with the double banana RG58 cable and BNC (M) to the binding post adapter.
  • Page 67 Calibrator Calibration Verification of the Product 4. Populate Table with the Shunt Phase Error. Obtain the phase error from the A40B specifications or phase calibration process. 5. Starting with step 8, Table 28, source the specified phase with the specific signal and frequency parameters from the Product.
  • Page 68: Frequency Calibration

    5560A/5550A/5540A Service Manual Table 28. Phase Calibration Record for V/I Shunt Measured Calculated Product Product Step Product Output Phase Product Product Limits Frequency Phase Error Output Output 10 V/0.1 A 65 Hz 0° ±0.10 ° 10 V/0.1 A 65 Hz 30°...
  • Page 69: Calibration Adjustment

    All the routine calibration steps are available from the front panel interface as well as the remote interface (IEEE-488 or serial). Remote commands for calibration are described in the 5560A/5550A/5540A Remote Programmers Manual located at flukecal.com.
  • Page 70: Start The Calibration Adjustment

    Volts DC Calibration Adjustment (OUTPUT VZ) The Product begins the adjust with internal steps.Complete a Vdc Zero on the Fluke Calibration 8588A. Put a 4-wire short (Fluke PN 2540973) across the HI and LO input and sense terminals.
  • Page 71: Volts Ac Calibration Adjustment (Output Vz)

    The Product begins the adjustment with internal steps. The volts ac calibration uses two methods. For frequencies ≤ 10 Hz, a low frequency DMM (Fluke Calibration 8588A) is used. For frequencies >10 Hz, use the Fluke Calibration 5790B AC Measurement Standard.
  • Page 72: Frequency Calibration Adjustment (Output Vz)

    5560A/5550A/5540A Service Manual Frequency Calibration Adjustment (OUTPUT VZ) The frequency calibration adjustment requires a frequency counter - see Table for the recommended model/requirements. 1. When the Product displays a connect prompt, connect the frequency counter Input A to the Product OUTPUT VZ HI/LO with the BNC cables and adapters 2.
  • Page 73: Volts Dc (Dcv Dcv) Calibration Adjustment (Secondary Output) (5550A And 5560A)

    The Product begins the adjustment with internal steps. Complete a Vdc Zero on the Fluke Calibration 8588A - put a 4-wire short (Fluke PN 2540973) across the HI and LO input and sense terminals. 1. Set the 8588A as follows: Volts dc, Range Auto, NPLC 50, Guard ON (external guard).
  • Page 74: Resistance Calibration Adjustment

    5560A/5550A/5540A Service Manual 4. Measure with the meter and input the value on the touch screen in the Enter Measured Value field. Note that the Product only accepts values in base units. Example: a measurement output is shown as 12.0000 mV, but the Product expects the value to be input in V.
  • Page 75 Calibrator Calibration Adjustment Table 31. Calibration Adjust Steps for Capacitance with LCR meter Step Product Output Test Frequency 0.2 nF 1000 Hz 1 nF 1000 Hz 1.21 nF 1000 Hz 10 nF 1000 Hz 12.1 nF 610 Hz 20 nF 610 Hz 30 nF 610 Hz...
  • Page 76: Inductance Calibration Adjustment

    5560A/5550A/5540A Service Manual 7. Set the LCR meter test frequency. 8. Push OPERATE to set the Product to Operate. 9. After the LCR meter measurement settles, input the measurement in the Enter Measurement Value window on the Product. 10. Complete steps 6 to 9 until you complete all points in the sequence.
  • Page 77: Tc Source Calibration Adjustment

    Calibrator Calibration Adjustment Table 32. Calibration Steps for Inductance with LCR Meter (cont.) Step Product Output Test Frequency 10 Hz 1.21 H 3 Hz 10 H 3 Hz 12.1 H 2 Hz 100 H 2 Hz 5. Push OPERATE to set the Product to Operate. 6.
  • Page 78: Tc Measure Calibration Adjustment

    5560A/5550A/5540A Service Manual TC Measure Calibration Adjustment The Product begins the adjustment with a prompt to connect to a vdc calibrator. 1. Connect the Type B copper mini-plug with copper wire between a calibrator and the Product TC INPUT/OUTPUT. Allow 3 minutes for thermal dissipation.
  • Page 79: Replace The Mains Power Fuse

    5. Push the fuse compartment door back into place until the tab locks. Table 33. Replacement Fuses Line Voltage Range Fuse Description Fluke Calibration Part Number 100 V - 120 V T 5.0 A 250 V 109215 220 V - 240 V T 2.5 A 250 V...
  • Page 80 5560A/5550A/5540A Service Manual Figure 29. Access the Fuse...
  • Page 81: Replaceable Parts

    (PCA) level should only be performed by a Fluke Calibration Authorized Service Center. Units found to be unable to adjust within specification or exhibit error codes during diagnostics indicate that the Product will require service and should be returned to a Fluke Calibration Authorized Service Center for troubleshooting and repair.
  • Page 82 5520A-3013,SHIM,TRANSFORMER 625985 SCREW,HK M4-0.7X8MM,STEEL,PLAIN,LOW HEAD 40 (not shown) 3472154 SOCKET CAP SCREW,W/SELF LOCKING PATCH 43 (not shown) 5560A-8007,NUT, #10, LOW THERMAL 4859582 WASHER,LOCK,EXT TOOTH,#10 SCREW,.204IN 44 (not shown) 4884555 ID,.395IN OD,.018IN THK,PHOS BRONZE WASHER,FLAT,#10 X.5IN OD,.04IN THK,GRADE C- 45 (not shown)
  • Page 83 SPRING,COMPRESSION,0.24IN OD,1.75IN L,0.02IN 81 (not shown) 5272855 WIRE,1.5LB/IN,MUSIC WIRE 82 (not shown) 5520A-2026,TRANSFORMER COVER, PAINTED 647138 83 (not shown) 5560A-8015,DISPLAY CABLE SHIELD/BRACKET 4962599 5522A-8002,RETAINER, ANALOG TOPCOVER 3472691 NUT,HEX,K-LOCK,6-32,.140IN THK,5/16IN AF,EXT 89 (not shown) 152819 TOOTH LOCK WASHER,STEEL 90 (not shown)
  • Page 84 97 (not shown) 5194811 OD,.043 THK,.068 H,STEEL,ZINC CLR FOAM PAD,URETHANE,.250 IN X.375 IN X .062 IN 98 (not shown) 2567386 THK,ADHESIVE 5560A-2027,XFRMR BRACE 5334517 5560A-2028,BRACE INSULATOR PLATE 5334521 MAGNETIC MATERIAL,ROUND CABLE EMI 101 (not SUPPRESSOR,265OHMS@100MHZ,12.7MM 5383876 shown) CABLE,30X40MM 102 (not...
  • Page 85 Calibrator Replaceable Parts Figure 30. Replacement Parts...
  • Page 86 5560A/5550A/5540A Service Manual...