Fluke 5500A Operator's Manual page 107

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

8. Press the numeric keys and decimal point key to enter the desired frequency output
(maximum five numeric keys). Press a multiplier key, if necessary. For example,
press the kilo multiplier key
1.1234 kHz (below).
9. Press
. The calibrator clears your entry from the Control Display and copies
it into the Output Display (below is typical).
10. Press
to activate the calibrator output.
Several softkey labels appear on the Control Display in the ac current function,
depending on the 5500A configuration: OUTPUT and WAVE.
OUTPUT (Output Location) Selects the current output terminals between the
5500A Calibrator (aux) and 5725A Amplifier (boost). If a 5725A Amplifier is not
connected, the OUTPUT softkey label will not appear in the display.
WAVE (Waveform) Allows you to select one of four different types of waveforms:
sinewave, trianglewave, squarewave, and truncated sinewave. (See "Waveform
Types" later in this chapter for more information). Whenever a non-sinusoidal
waveform is selected, the Output Display will convert the RMS reading to peak-to-
peak (P
).
P
. Then press the
key. For example,
Front Panel Operation
Setting the Output
4
4-23