Loop Power Supply
Voltage: 24 V
Maximum current: 22 mA
Short circuit protected
Pressure Measurement (725 and 726)
Range
Resolution
Determined by
4 digits
pressure module
Pulse Read and Pulse Source (726)
Pulse
Source
1
Read
Pressure Units Availability (725)
Unit
psi
inH
O
2
cmH
O
2
bar
mbar
kPa
inHg
mmHg
2
kg/cm
Accuracy
Determined by
pressure module
Min
Max
10,000
100,000
Available on all pressure ranges
All ranges through 3000 psi
All ranges through 1000 psi
15 psi and above
All ranges through 1000 psi
Available on all pressure ranges
Available on all pressure ranges
All ranges through 1000 psi
15 psi and above
Units
Pushing for 3
psi, inH
O@4 °C,
2
inH
O@20 °C, kPa,
seconds stores
2
cmH
O@4 °C,
present pressure
2
cmH
O@20 °C, bar,
value as an offset and
2
mbar, kg/cm 2,
subtracts it from the
mmHg, inHg
displayed value.
Accuracy
1 Count
Availability
Calibrators
Specifications
Mode (726 Only)
Frequency
2 CPM to 10 kHz