Time Marker Function Specifications; Wave Generator Specifications - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

8-81. Time Marker Function Specifications

Time Marker into 50
1-Year Absolute
Uncertainty, tcal
5 C [3]
Wave Shape
Typical Output level
Sequence (cardinal
points)
Adjustment Range
Resolution
[1] t is the time in seconds. Examples: At 5 s the uncertainty is 5,025 ppm; At 50 s the uncertainty is
25.75 ppm.
[2] The 2 ns time marker is typically > 0.5 V p-p.
[3] Away from the cardinal points, add 50 ppm to uncertainty.

8-82. Wave Generator Specifications

Wave Generator Characteristics
Amplitude
Range
1-Year Absolute Uncertainty, tcal
10 Hz to 10 kHz
Sequence
Typical DC Offset Range
Frequency
Range
Resolution
1-Year Absolute Uncertainty, tcal
[1] The DC offset plus the wave signal must not exceed 30 V rms.
5 s to 100 s
50 s to 2 s
(25 + t*1000)
(25 + t*15,000)
ppm [1]
ppm [1]
pulsed
pulsed
sawtooth
sawtooth
> 1 V pk
> 1 V pk
5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms)
At least
Square Wave, Sine Wave, and Triangle Wave
into 1 M :
into 50
5 C,
± (3% of p-p output + 100 µV)
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
0 to ± ( 40% of p-p amplitude) [1]
10 Hz to 100 kHz
4 or 5 digits depending upon frequency
5 C
Oscilloscope Calibration Option Specifications
1 s to 20 ns
25 ppm
pulsed
sawtooth
> 1 V pk
10% around each cardinal point.
4 digits
into 50
or 1 M
1.8 mV to 55V p-p
:
1.8 mV to 2.2V p-p
(25 ppm + 15 mHz)
5500A-SC300 Option
10 ns to 2 ns
25 ppm
sine
> 2 V p-p [2]
8
8-63

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents