8-50. General Commands

SCOPE
(IEEE-488, RS-232, Sequential)
Programs the 5520A-SC oscilloscope calibration hardware, if installed. The instrument
settings are determined by this command's parameter. Once in SCOPE mode, use the
OUT command to program new output in all functions except Impedance Measurement
and the RANGE command as required (in OVERLD, PULSE, and MEAS Z functions
only.) OPER, STBY, *OPC, *OPC?, and *WAI all operate as described in Chapter 6.
The state of the oscilloscope's output while in SCOPE mode is reflected by the bit in the
ISR that is assigned to SETTLED.
Parameter
OFF
VOLT
EDGE
LEVSINE
MARKER
WAVEGEN
Table 8-15. SCOPE Command Parameters
Turns the oscilloscope hardware off. Programs 0 V, 0 Hz, output at the
NORMAL terminals, standby.
Oscilloscope ac and dc VOLT mode. Programs 20 mV peak-to-peak, 1 kHz,
output at the SCOPE BNC, output impedance 1 M , standby if from OFF or
previously in standby. FUNC? returns SACV (for ac) or SDCV (for dc).
Example:
SCOPE VOLT; OUT 4 V, 1 kHz
(ac voltage, 4 V peak-to-peak, 1 kHz.)
Oscilloscope EDGE mode. Programs 25 mV peak-to-peak, 1 MHz, output at the
SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns
EDGE.
Example:
SCOPE EDGE; OUT 0.5 V, 5 kHz
(Edge, 0.5 V peak-to-peak, 5 kHz.)
Oscilloscope LEVSINE mode. Programs 30 mV peak-to-peak, 50 kHz, output at
the SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns
LEVSINE.
Example:
SCOPE LEVSINE; OUT 1 V, 50 kHz
(Leveled sine wave, 1 V peak-to-peak, 50 kHz.)
Oscilloscope MARKER mode. Programs the period to 1 ms, output at the
SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns
MARKER.
Example:
SCOPE MARKER; OUT 2 MS
(Marker, period of 2 ms.)
Oscilloscope WAVEGEN mode. Programs 20 mV peak-to-peak, square wave,
1 kHz, no offset, output impedance 1 M , standby if from OFF or previously in
standby. FUNC? returns WAVEGEN.
Example:
SCOPE WAVEGEN; OUT 1 V, 1 kHz
(Wave Generator, 1 V peak-to-peak, 1 kHz.)
Remote Commands and Queries
Description/Example
5500A-SC600 Option
8