Fluke 5500A Operator's Manual page 224

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual
Each example shows a value and unit, e.g., -15.2 V. If a value is entered without a unit,
the value of the existing output is changed, when logically allowed.
OUT?
Returns the output amplitudes and frequency of the 5500A Calibrator. Multipliers (e.g.,
K or M) are not used in the response.
Parameters: V
Response:
Examples:
The respective values for the above examples are:
6-28
OUT 10 V, 1 A
OUT 15 DBM, .5 A
OUT 1 V, 1 A, 60 HZ
OUT 5 DBM, 1 A, 50 HZ
OUT 1 V, 2 V
OUT 8 DBM, 12 DBM
OUT 10 MV, 20 MV, 60 HZ (dual volts; .01 V, .02 V @ 60 Hz)
OUT 6 DBM, 8 DBM, 50 HZ (dual volts; 6 & 8 dBm @ 50 Hz)
x
x
IEEE-488
RS-232
(optional for ac voltage and TC modes)
(optional for ac voltage modes)
DBM
(optional for RTD and TC modes, Celsius)
CEL
(optional for RTD and TC modes, Fahrenheit)
FAR
(optional for RTD modes, ohms)
OHM
<primary amplitude value>,<primary units>,
<secondary amplitude value>,<secondary units>,
<fundamental frequency value>
OUT? returns -1.520000E+01,V,0E+00,0,0.00E+00
OUT? returns 1.88300E-01,A,0E+00,0,4.420E+02
OUT? returns 1.23000E+00,V,2.34000E+00,V,6.000E+01
OUT? DBM, DBM returns 4.02E+00,DBM,9.60E+00,DBM,6.000E+01
OUT? DBM, V returns 4.02E+00,DBM,2.34000E+00,V,6.000E+01
OUT? returns 1.92400E+06,OHM,0E+00,0,0.00E+00
OUT? returns 1.52000E+01,V,1.88300E-01,A,4.420E+02
OUT? DBM returns 2.586E+01,DBM,1.88300E-01,A,4.420E+02
OUT? returns 1.0430E+02,CEL,0E+00,0,0.00E+00
OUT? FAR returns 2.19740000E+02,FAR,0E+00,0,0.00E+00
OUT? V returns 4.2740E-03,V,0E+00,0,0.00E+00
OUT? OHM returns 1.40135E+02,OHM,0E+00,0,0.00E+00
-15.2 V
188.3 mA, 442 Hz
1.23 V, 2.34 V, 60 Hz
4.02 dBm, 9.6 dBm, 60 Hz (4.02 dBm=1.23 V; 9.6 dBm = 2.34 V)
4.02 dBm, 2.34 V, 60 Hz (4.02 dBm=1.23 V)
1.924 M
15.2 V, 188.3 mA, 442 Hz
25.86 dBm, 188.3 mA, 442 Hz (25.86 dBm = 15.2 V)
104.3 C
(power; 10 watts @ same frequency)
(power; 7.5 watts @ same frequency)
(power ac; 1 watts @ 60 Hz)
(power ac; 5 watts @ 50 Hz)
(dual volts; 1 V, 2 V @ same freq.)
(dual volts ac; 8/12 dBm @ same freq.)
x
x
Sequential
Overlapped
x
Coupled

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents