Fluke 5500A Operator's Manual page 214

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual
Temperature Standard
Host Connection
GPIB Port Address
Serial Ports
EOL (end of line)
EOF (end of file)
Remote I/F
SRQSTR
* Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6,7.
FUNC?
(Function query) Return the present output, measurement, or calibration function. See
the response below for output & measurement modes.
Responses:
Example:
Return DCV_DCV when the 5500A Calibrator output function dual dc volts.
6-18
its-90
gpib (IEEE-488)
4
8 bits, 1 stop bit, xon/xoff, parity
none, 9600 baud, wait 30 sec.
CRLF
012,000
term
Remote commands (see Chapter 6)
SRQ: %02x %02x %04x %04x
x
x
IEEE-488
RS-232
(dc volts function)
DCV
(ac volts function)
ACV
(dc current function)
DCI
(ac current function)
ACI
(ohms function)
RES
(capacitance function)
CAP
(temperature with an rtd function)
RTD
(temperature with a thermocouple function)
TC_OUT
(dc power function)
DC_POWER
(ac power function)
AC_POWER
(dual dc volts function)
DCV_DCV
(dual ac volts function)
ACV_ACV
(measure temperature with a thermocouple)
TC_MEAS
SACV
(oscilloscope ac volts function)
SDCV
(oscilloscope dc volts function)
MARKER
(oscilloscope marker function)
LEVSINE
(oscilloscope leveled sine function)
EDGE
(oscilloscope edge function)
FUNC? returns DCV_DCV
Features
Display Contrast*
Display Brightness*
RTD Power Up
Default Type
Thermocouple Power
Up Default Type
Source Preference
Current Limits
Voltage Limits
*PUD
x
x
Sequential
Overlapped
level 7,7
level 1,0
pt385
K
5500
+11 A
+1000 V
cleared
string
x
Coupled

Hide quick links: