7-25. AC Voltage Amplitude Accuracy, Squarewave (NORMAL)

This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The AC Voltage Amplitude Accuracy, Squarewave (NORMAL) tests the amplitude
accuracy at the NORMAL terminals. For this test, use the Fluke 5790A, an rms-
responding meter. For squarewaves, the measured value (in rms) should be exactly 1/2
the nominal value in peak-to-peak.
Nominal Value (V p-p)
30 mV (15 mV rms)
30 mV
30 mV
30 mV
300 mV (150 mV rms)
300 mV
300 mV
300 mV
3 V (1.5V rms)
3 V
3 V
3 V
30 V (15 V rms)
30 V
30 V
30 V
7-22
Note
Frequency (Hz)
Measured Value
10 Hz
1 kHz
20 kHz
100 kHz
10 Hz
1 kHz
20 kHz
100 kHz
10 Hz
1 kHz
20 kHz
100 kHz
10 Hz
1 kHz
20 kHz
100 kHz
Deviation
(V rms)
%
1-Year Spec.
(%)
1.350%
0.800%
1.050%
6.100%
1.350%
0.800%
1.050%
6.100%
1.350%
0.800%
1.050%
6.100%
1.350%
0.800%
1.050%
6.100%