5500A
Operator Manual
8-14. Oscilloscope Input Resistance Measurement Specifications
Scope input selected
Measurement Range
Uncertainty
8-15. Oscilloscope Input Capacitance Measurement Specifications
Scope input selected
Measurement Range
Uncertainty
[1] Measurement made within 30 minutes of capacitance zero reference. Scope option must be selected
for at least five minutes prior to any capacitance measurement, including the zero process.
8-16. Overload Measurement Specifications
Source
Voltage
5 V to 9 V
8-12
Table 8-12. Oscilloscope Input Resistance Measurement Specifications
40
Table 8-13. Oscilloscope Input Capacitance Measurement Specifications
Table 8-14. Overload Measurement Specifications
Typical 'On' current
indication
100 mA to 180 mA
50
to 60
0.1 %
1 M
5 pF to 50 pF
(5% of input + 0.5 pF) [1]
Typical 'Off' current
indication
10 mA
1 M
500 k
to 5 M
0.1 %
Maximum Time Limit DC
or AC (1 kHz)
setable 1 s to 60 s