Fluke 5500A Operator's Manual page 139

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

8. Test the Ohms function as follows:
a. Press
on the calibrator and set the DMM function switch to
b. Set the calibrator to 190.0
. Verify the error is within specifications.
c. Repeat the previous step for 19.00 k , 1.900 M , and 19.00 M . Verify the
errors are within specifications.
d. Press
conductance tests of high resistances.
e. Set the calibrator output to 100 M . Verify the error is within specification.
9. Test the capacitance function as follows (use the REL feature of the 80 Series to
subtract cable capacitance):
a. Press
on the calibrator and set the DMM function switch to
press the blue button.
b. Set the calibrator output to 1.0 F with compensation off. Press
error is within specification.
c. Repeat the previous step using 0.470 F, 0.047 F, and 4.70 nF. Verify the errors
are within specifications.
10. Test the Diode Test function as follows:
a. Press
on the calibrator and set the DMM function switch to
b. Set the calibrator to 3.0V dc and press
specification.
11. Test the ac and dc current function:
a. Press
on the calibrator and set the DMM function switch to
b. Verify that the calibrator is in standby and connect the DMM as shown in
Figure 4-16.
87
TRUE RMS MULTIMETER
MIN MAX
RANGE
HOLD
H
REL
Hz
PEAK MIN MAX
mV
mA
V
A
A
V
OFF
mA A
COM
A
V
Figure 4-16. Cable Connections for Testing an 80 Series Current Function
with 2-wire compensation (see Figure 4-2). Press
on the DMM to enter the 40-nanosiemen range, used for
Front Panel Operation
. Verify the error is within
5500A
NORMAL AUX
V,
,
RTD
HI
1000V
RMS
MAX
LO
1V PK
MAX
20V PK
MAX
TC
4
Sample Applications
.
and
. Verify the
.
.
CALIBRATOR
SCOPE
200V PK
A,
-SENSE,
MAX
AUX V
20V
RMS
MAX
TRIG
OUT
20V PK
MAX
f4-16.eps
4-55

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents