System description
Furthermore it is necessary that the control checks all status signals of the E20, also the 0-sheet
signal. While measuring, only the 1-sheet signal may occur. If 0-sheet or 2-sheet occurs, the
measured sheet may not be transported on. In such cases, the sheet will usually be deposited, re-
picked and measured again.
The control needs to evaluate all status signals!
Checking-up through the 0-sheet signal is to assure that a sheet really is in front of the sensor
during measuring. If 0-sheet occurs while measuring, the measurement is not correct. For example,
a defective sensor holder might prevent the sensor from contacting the sheet. Monitoring the sheet
thickness would then only be performed apparently.
In case of double sheet fatal consequences could then result.
Pos: 24 /Doppelblech/Geräte/E20 Bus/3 Systembeschreibung/Auswertegerät @ 0\mod_1173874473250_501.docx @ 3860 @ 2 @ 1
3.3

Control unit

The E20 is based on the product platform R1000.
Major features of the control unit:
Programmable for 255 different sheet thicknesses
Fieldbus interface for the control, operation and data exchange with the PLC
Calibration by a teach-in procedure (see "3.3 Calibration")
Digital display of sheet thickness and operating parameters
Monitoring of over gauge and under gauge limits
Monitoring of operating voltage and sensor gap
Optocoupler for direct I/O control via the PLC for fast measurement operations
Software firmware update
Data backup via fieldbus interface or serial interface
Pos: 25 /Doppelblech/Geräte/E20 Bus/3 Systembeschreibung/Parameter des Auswertegerätes @ 1\mod_1195739891578_501.docx @ 5239 @ 233 @ 1
3.4

Parameters of the control unit

The E20 provides for extensive configuration. So, different customer requirements can be served.
Configuration is performed by setting system parameters and program parameters.
System parameters are set only once on commissioning, and will then remain unchanged. Program
parameters contain items which are material and job depending, and enable individual adaptation
to the measuring task.