3053-3
ML700 from SN 1023 AND HIGHER AND ALL 60Hz UNITS
1
.
Connect negative lead of DVM to TP1 of PCB A7.
2
.
Connect positive lead of DVM to TP 2 of PCB A7.
3
Adjust R5 on PCB A7 to 7.5V ± 75 mV.
.
4
.
Connect positive lead of DVM to TP 3 of PCB A7.
5
.
Cover SENSOR B14 with DENSITY TOOL PN 9191223
(ND 1.5 WRATTEN FILTER #96)
6
.
Adjust R2 on PCB A7 to 5V.
7
.
Adjust the mechanical position of the DOUBLE FILM SENSOR until the DVM shows
the highest reading.
8
Adjust R2 on PCB A7 to 7.7 V ± 0.1V .
.
9
.
Test with fresh CUSTOMER FILMS (Single and double).