8-6. Correcting The Internal Mass Value Of The GX-K Series

The balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected
mass value is maintained in non-volatile memory even if the AC adapter is removed. The
internal mass value is corrected as follows:
Note GF-K series can not use this function.
Model
GX-8K GX-8K2
GX-10K
GX-12K GX-20K
GX-30K GX-32K
Step 1 Calibrate the balance using the internal mass. (one-touch calibration).
Place the external weight and find out the correction value.
Example for correcting the weight value by +0.3 g in 10 kg, when
using GX-12K. If correcting the weight value by +0.3 g in 5 kg,
the weight changed into 10 kg, the correction value is +0.6 g.
Step 2 Press the
Step 3 While pressing and holding the
the
SAMPLE
balance displays
Step 4 Press the
switches. Set the function table switch and internal mass
correction switch to "
SAMPLE
RE-ZERO
Step 5 Press the
The balance returns to the weighing mode.
Step 6 Press and hold the
table and release the key when
Step 7 Press the
displayed, then release the key.
Step 8 Press the
correcting the internal mass value.
Step 9 Correct the internal mass value using the following keys.
RE-ZERO
key ....... To store the new value and display the
PRINT
Target Range
2kg
±0.50g
10kg
±5.0g
key to turn off the display.
ON:OFF
key, press the
.
p5
key. Then the balance displays the function
PRINT
" as shown above using the following keys.
1
key ... To select the switch to change the value.
The selected digit blinks.
key .. To change the parameter of the switch selected.
Function table switch
Correction switch of internal mass
key to store the new setting.
PRINT
key to enter the function
SAMPLE
key several times until
SAMPLE
key to enter the procedure for
PRINT
key .. To select the value. (-50 digits appear
after +50 digits)
next menu item of the function table.
10kg
Correction the internal
mass by +0.3g at 10kg.
Calibrate with the
10000. 0 g
corrected internal mass.
Corrected external weight
key and
PRINT
key. The
ON:OFF
is displayed.
ba5fnc
C5 in
29
The same mass
With these keys Press the key
held down
Press and hold
is
10kg
10000. 3 g