Exchanging the Seal Wash Pump Cartridge

When
If mechanically defect
Parts required
#
1
1 Disconnect seal wash tubing at the connectors, and place the ends of each
2 By pressing the upper and lower clip remove the seal wash pump cartridge.
3 Replace the defective cassette and slide the new cassette on to the shaft
4 Reconnect seal wash tubing.
Agilent 1260 Infinity Preparative Pump User Manual
Part number
Description
5042-8507

Seal wash pump cartridge

tubing at a level above seal wash bottle (to avoid leaking solvents).
pressing upper and lower clip. Push forward until the clips snap into place.
The preferred orientation of the cartridge is as shown in picture below.
Figure 29
Seal Wash Pump Cartridge
7
Maintenance
Simple Repairs
125