Calibrating Sensor Head

Calibrating Sensor Head p.66
Operating procedure
Steps
Key operation
1
2
3
4
5
Important
When Sensor Head calibration fails
If the Sensor Head is calibrated in an inappropriate environment, an error is displayed on the main display. In such a
case, return to the previous screen using
If an error results even if Sensor Head calibration is done under the correct conditions, it is possible that some
contaminants may have adhered to the Sensor Head part, or the fiber connector of the Sensor Controller. Clean the
fiber cable or fiber connector referring to p.59.
244
Setting the System
Display
Description
Press and hold the
mode.
(For details on the functions of the Mode switching key, see p.197.)
Press the
Press the
"OK/CAN" is displayed on the sub-display.
Press the
Press the
key. Then refer to the p.66, recheck your environment, and try again.
key for two seconds to enter the FUNC
/
keys to select "SYSTEM" and press the
/
keys to select "H.CALIB", then press the
key to execute calibration or the
key for two seconds to enter the RUN mode.
key.
key.
key to cancel.