Performing Internal Logging with Parallel I/O

The logging start and end timings can be specified using LOGGING input signals from parallel I/O. When doing
so, follow the procedure below to set the data logging conditions.
 Multi view Explore
→ Edit pane
→ Sensor settings window
1
Set the logging conditions for internal logging.
There are three settings: Number of logging, number of logging decimation, and overwrite mode.
Item
Internal logging
setting
LOGERR signal is turned ON at the following cases:
• Overwrite mode is OFF
• 2,000,000 items have already logged in the internal memory at the begging of logging.
When continue to logging regardless number of logging while LOGGING input signal is turned ON, set
as below:
• Number of logging data: 0
• Overwrite mode: ON
108
Performing internal logging
Internal
1
0.2
OUT1
-1938
-1939
-1940
-1940
-1941
-1941
: [System] (double-click)
: [Sensor Settings] icon (
: [Internal logging setting]
Setting item
Setting value
Number of
0 to 2000000
logging data
(default value: 100000)
Internal logging
0 to 1000
settings
(default value: 1)
Overwrite mode
OFF (default value)/ON
Label insert
OFF (default value)/ON
mode
OUT2
OUT3
2823
9437
2764
9401
2773
9355
2807
9523
2794
9442
2817
9497
)
Description
Set the number of data to save for each internal logging
output data.
Set the interval of saving internal logging data.
If 1 is set, all data is saved. If 2 is set, one measured data is
saved every two measurements.
If 0 is set, data is saved only when the hold value is applied.
Set whether to make the internal logging mode run in
overwrite mode.
Sets the operation to be performed when the logging count
exceeds the set number of logging data.
OFF: Internal logging is stopped.
ON: The oldest logging data is overwritten.
Set the operation to be performed when logging input is
turned on.
OFF: No operation.
ON: When logging input is turned on, inserts label into log-
ging data.
OUT4
-2147483648
-2147483648
-2147483648
-2147483648
-2147483648
-2147483648
ZW-7000/5000
User's Manual