Omron ZW-7000 Series User Manual page 143

Confocal fiber type displacement sensor
Hide thumbs Also See for ZW-7000 Series:
Table of Contents

Advertisement

 Multi View Explore
→ Edit pane
→ Task Settings window : [Filter]
1
Select the frequency filter type from
[Frequency filter].
2
Set the cutoff frequency from [Cutoff
frequency]. When a band pass filter is
selected, set [Cutoff frequency (upper limit)]
and [Cutoff frequency (lower limit)].
The range of cut-off frequency to be effective in operation varies depending on the setting of measurement
cycle.
The relationship between the measurement cycle and the cut-off frequency range is shown in the table below.
Measurement cycle
Sampling
[μs]
Frequency
[Hz]
20
50000
40
25000
60
16667
80
12500
100
10000
200
5000
300
3333
400
2500
Note
The frequency filter can also be set by the operating keys on the Sensor Controller.
Setting the Frequency Filter p.219
Important
• If a value outside of the appropriate range is set, the frequency filter will be applied with the values included in the
range.
• If the cut-off frequency values (upper/lower limits) are too close, the frequency bandwidth to be detected cannot be
properly detected.
• If changing the extension fiber cable length setting causes the setting value of the cutoff frequency to fall outside the
upper/lower limit value, the setting value is automatically changed to the value within the limit.
ZW-7000/5000
User's Manual
: [Bank] | [(Bank Data Name)] (double click)
: [Task Settings] icon (
High-pass filter
Band-pass filter
Lower limit
Upper limit
1275
23725
638
11862
425
7907
319
5931
255
4744
128
2372
85
1581
64
1185
)
Cut-off frequency range [Hz]
Low-pass filter
Lower limit
20
10
7
5
4
2
1.5
1
Setting the Output Conditions
Upper limit
23725
11862
7907
5931
4744
2372
1581
1185
141

Advertisement

Table of Contents
loading

This manual is also suitable for: