Bank Data
ZW-7000/5000 Series can hold up to eight sets of sensing settings, which are called "bank (BANK)".
When the setup is changed, the bank can be switched externally.
What is Bank?
The sensing settings for measurement
Sensing
Filter
Output
I/O
Example of switching bank for settings
If you register settings of various classes,
Sensing setting for product class A
Sensing
Filter
Output
I/O
Measuring operation
with the contents set
in BANK1
Sensing
Measurement point
Important
• If you want to register multiple bank data having the same values set except for "threshold values (judgment
values)", by changing the mode of the bank, you can increase the number of banks from 8 to 32.
Setting the Bank Mode p.168
• The "output destination" (current output value/voltage output value) in the I/O setting parameters is set the same for
all banks. The output destination cannot be set separately for individual banks.
"2-1 Parallel I/O connection" described in Displacement Sensor ZW-7000 series Confocal Fiber Type
Displacement Sensor User's Manual for Communications Settings (Z363)
ZW-7000/5000
User's Manual
Threshold value
Register the different settings
for changing the device setup.
Threshold value
BANK1
Switch
to BANK2.
BANK2
BANK3
BANK8
TASK1
Filter
Threshold value
Output
I/O
are held as one bank.
Sensing
Filter
Output
I/O
Sensing setting for product class B
Sensing
Filter
Output
It is easy to change the device setup.
TASK4
Measurement point
Filter
Threshold value
Output
BANK1
Threshold value
BANK
Measuring operation
with the contents set in
BANK2
I/O
Threshold value
BANK1
BANK2
BANK3
BANK8
TASK1 to TASK4 are
specified in the settings
for each bank.
Bank switching