Mate the convex section of the fiber cable with the groove on the fiber connector and turn the
threaded section clockwise while pushing in (see figure 1).
(figure1)
Fiber connector
In the "No good" status in figure 2 and figure 3, optic signals cannot be transmitted and correct measure-
ment is not possible. Always check that the system is in the "Good" status.
3
Connect the calibration ROM to the Sensor Control-
ler, and then tighten the supplied M2 screw.
Tightening torque: 0.15 N · m max.
Important
• Before connecting or disconnecting the calibration ROM, make sure that the Sensor Controller's power supply is
turned OFF. The Sensor Controller may break down if the calibration ROM is connected while the power is ON.
• Use with the Calibration ROM always connected. If the Calibration ROM is not connected, the error message
"SYSERR" is shown in the main digital and "NO.ROM" is shown in the sub digital, where measurement is not possi-
ble.
If the Calibration ROM is lost, or fails, refer to p.174.
"6-1 Error Messages" described in Displacement Sensor ZW-7000 series Confocal Fiber Type Dis-
placement Sensor User's Manual for Communications Settings (Z363)
• Be sure to use a Sensor Head and Calibration ROM with the same serial number. A pair with different serial num-
bers cannot operate normally.
• Make sure to initialize the Sensor Controller when connecting a different type Sensor Head which is used at the last
launch to the Sensor Controller.
ZW-7000/5000
User's Manual
Groove
section
Convex
section
Threaded
section
Fiber adapter
(figure2)
(figure3)
M2 screw
Calibration ROM
ROM connector
57