Omron ZW-7000 Series User Manual page 263

Confocal fiber type displacement sensor
Hide thumbs Also See for ZW-7000 Series:
Table of Contents

Advertisement

l ZW-S5010/S5020/S5030
Item
Sensor controller
Measurement center distance
Measuring range
*1
Static resolution
*2
Linearity
Spot diameter (Total measurent
*3
range)
*4
Measuring cycle
Operating ambient illumination
Ambient temperature range
Ambient humidity range
Degree of protection
Vibration resistance (destructive)
Shock resistance (destructive)
*5
Temperature characteristic
LED Safety
Material
Fiber cable length
Fiber cable minimum bend radius
Insulation resistance (Calibration
ROM)
Dielectric strength (Calibration
ROM)
Weight
Accessories
*1:
Capacity value when OMRON standard mirror surface target is measured at the measurement center distance as the average of 16,384
times
When connected with the controller for the Export Control Trade Ordinance (ZW-5000T), the minimum resolution is 0.25
the Sensor head or setting condition.
*2:
Material setting for the OMRON standard mirror surface target: Error from an ideal straight line when measuring on mirror surface.
*3:
Capacity value defined by 1/e
*4:
When an extension fiber cable of 5 mm or longer is connected, the setting rage of the measurement cycle (exposure time) changes. For
details, refer to "Setting Measurement Cycle" in this manual.
*5:
Capacity value of temperature characteristic at the measurement center distance when fastened with an aluminum jig between the Sensor
Head and the target and the Sensor Head and the Sensor Controller are set in the same temperature environment.
ZW-7000/5000
User's Manual
Specifications
ZW-S5010
ZW-5000@
10 mm
±0.5 mm
0.004 μm
±0.45 μm
9 μm dia.
80 μs to 1600 μs
Illumination on object surface max.30000 Lx: (incandescent light)
Operation: 0 to 50°C, Storage: -15 to +60°C
(No freezing and condensation)
Operation/storage: 35 or 85%RH (No condensation)
IP40 (IEC60529)
10 to 150 Hz (half amplitude 0.35 mm), 80 mins in each of X/Y/Z directions
2
150 m/s
, 6 direction, 3 times each (up/down, left/right, forward/backward)
0.6 μm/°C
Risk Group 3 (IEC62471)
Chassis: aluminum die cast
Fiber cable sheath: PVC
Calibration ROM: PC
0.3 m, 2 m (flex-resistant cable)
20 mm
Between case and all terminals: 20 MΩ (by 250 VDC)
Between case and all terminals: 1000 VAC, 50/60 Hz, 1 min
Fiber cable length 0.3m Approx. 170g
Fiber cable length 2m Approx. 180g
Calibration ROM fixing screws (M2), Fiber cable protective cap, Strap (1), Instruction Manual,
Precautions
2
(13.5%) of the peak optical intensity of the measurement wavelength.
ZW-S5020
20 mm
±1 mm
0.008 μm
±0.9 μm
13 μm dia.
1.1 μm/°C
Specifications and External Dimensions
ZW-S5030
30 mm
±2 mm
0.016 μm
±2.0 μm
18 μm dia.
1.8 μm/°C
μ
m, regardless of
261

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zw-5000 series

Table of Contents