4-4 Setting Threshold Value

Threshold Value Settings

Switch the Sensor Controller to the FUNC mode and set the range in order for the measured value to be judged
as PASS.
Both HIGH and LOW threshold values are set.
As a judgment result, HIGH, PASS or LOW is output.
HIGH threshold value
Measured value
LOW threshold value
ON
HIGH
OFF
ON
Output
PASS
OFF
ON
LOW
OFF
A judgment result appears on the Sensor Controller as follows.
• When the judgment result is "HIGH": HIGH indicator lights up
• When the judgment result is "PASS": PASS indicator lights up
• When the judgment result is "LOW": LOW indicator lights up
ZW-7000/5000
User's Manual
ON when Measurement result
> HIGH threshold value
ON when LOW threshold value
< Measurement result
< HIGH threshold value
ON when Measurement result
<LOW threshold value