Installation of the ADFM transducer assembly must also comply with the
following parameters to maintain the accuracy of the final installation:
• Transducer must be installed in such a way that the vertical beam
• Transducer must not be rotated about the vertical Z-axis. A sight-
• Transducer must be installed such that positive longitudinal flow
Positive
(Downstream)
direction of flow
+z out of page
+x
Figure 2-3.
2-5.3
Install Software and Configure Station File
Before the ADFM can be used in a specific application, it must be
programmed for that application. This is most easily done using the
WinADFM software. See
details.
page 2-8
(beam 5) is oriented normal with respect to the pipe's or chan-
nel's invert. Note that the vertical beam is normal to the trans-
ducer's top surface.
ing device or similar should be used to minimize rotational mis-
alignment.
direction is from the "non-cable" side of the transducer to the
"cable" side. See Figure 2-38 below for further explanation of
the flow direction.
+v
Transducer Orientation
Chapter 3
TOP
and the
Software Manual
MGD Technologies, Inc.
for further