5-4.3 Transmit Test - Teledyne ADFM Pro20 Technical Manual

Acoustic doppler flow meter
Hide thumbs Also See for ADFM Pro20:
Table of Contents

Advertisement

>pt3
Correlation Magnitude:
High Gain RSSI:
DAC Sin:
DAC Cos:
Duty:
Receive Test Results = $00000000 ... PASS
g. Observe the High Gain RSSI values. They should be between 20 to
25 counts with the transducer disconnected.
h. Leave the transducer cable disconnected if you want to run the next
test.
5-4.3
Transmit Test
This test runs a through test on the ADFM's transmit path electronic cir-
cuits.
a. Start BBTALK. Press the
b. Type
c. Type
>pt4
----------- BEAM 1 -----------
IXMT
=
VXMT
=
RXMT
=
Transmit Test Results = $40 ... PASS
----------- BEAM 2 -----------
IXMT
=
VXMT
=
RXMT
=
Transmit Test Results = $40 ... PASS
----------- BEAM 3 -----------
IXMT
=
VXMT
=
RXMT
=
Transmit Test Results = $40 ... PASS
----------- BEAM 4 -----------
IXMT
=
VXMT
=
ADFM Technical Manual (January 2000)
Lag
Bm1
Bm2
Bm3
0
255
255
255
1
192
196
198
2
70
75
3
9
11
4
16
15
5
12
15
6
10
11
7
9
8
21
21
186
187
59
50,
LPF:
End
CP
to clear the fault log.
PT4
. A message similar to the following should appear.
3.1 Ml Amps peak
4.1 Volts
peak
1335.6 Ohms
3.1 Ml Amps peak
4.1 Volts
peak
1335.6 Ohms
3.1 Ml Amps peak
4.1 Volts
peak
1335.6 Ohms
3.1 Ml Amps peak
4.1 Volts peak
Bm4
Bm5
255
255
192
194
88
76
73
27
26
11
10
26
17
6
9
13
10
15
8
10
19
7
20
21
21
0
key to wake the ADFM.
ADFM Test Procedures
page 5-5