CF - Flow Control:

Purpose:
Format:
Range:
Description:
Table C-3:
Command
CF1xxxx
CF0xxxx
CFx1xxx
CFx0xxx
CFxx1xx
CFxx0xx
CFxx1x
CFxx0x
CFxxx1
CFxxx0
ADFM Technical Manual (January 2000)
Selects when/how/where/what water-current data to collect.
CFnnnnn
Firmware switches (see description)
CF defines whether the ADFM: generates data ensembles
automatically or manually; generates pings immediately or
manually; sends output data in binary or HexAscii format;
sends or does not send output data out the serial interface;
sends or does not send output data to the recorder (if in-
stalled); retrieves or does not retrieve data from the recorder.
CMD Description
Description
Automatic Ensemble Cycling – Automatically starts the next data
collection cycle after data transmission is completed. Only a
<BREAK> can stop this cycling
Manual Ensemble Cycling - Enters the STANDBY mode after
transmission of the data ensemble, displays the > prompt, and
waits for a new command
Automatic Ping Cycling - Pings immediately when ready.
Manual Ping Cycling - Sends a < character to signal ready to ping,
and then waits to receive an <Enter> before pinging. The <Enter>
sent to the ADFM is not echoed. This feature lets you manually
control ping timing within the ensemble
Binary Data Output - Sends the ensemble in binary format.
HexAscii Data Output - Sends the ensemble in readable
hexadecimal-ASCII format.
Enable Serial Output - Sends the data ensemble out the RS-
232/422 serial interface.
Disable Serial Output - No ensemble data are sent out the RS-
232/422 interface.
Enable Data Recorder - Records data ensembles on the recorder
(if installed). The RD-command also must be set to zero to enable
the recorder.
Disable Data Recorder - No data ensembles are recorded on the
recorder.
Example:
CF1010 selects manual ensemble cycling, automatic ping cycling,
HexAscii data output, enables serial output, and disables data
recording.