3-4.3
Wakeup the ADFM
Wakeup is the process of establishing communication with the ADFM.
Immediately after you apply power to the ADFM, it enters the STANDBY
mode. Sending a BREAK signal from a terminal/program awakens the
ADFM. When the ADFM receives a BREAK signal, it responds with the
wake-up message. The ADFM is now ready to accept commands at the ">"
prompt from either a terminal or computer program. Press
wakeup command (BREAK) to the ADFM.
The ADFM wake-up (copyright) message should appear.
Broadband ADFM Version x.xx
RD Instruments (c) 1996-1998
All rights reserved.
>
If there is no response from the ADFM, check the communications setup in
the
SETUP MENU
NOTE. If the Caps Lock, Num Lock, or Scroll Lock keys are activated, this
may interfere with the ability to send a valid Break signal when pushing the
END key.
3-4.4
Communication Parameters
This menu sets the communications protocol between the ADFM and the
computer.
a. Press
b. Use the
c. Use the
d. Press
e. To permanently save this setup to the BBTALK.PTR configuration
file, press
BBTALK resides.
Figure 3-3.
ADFM Technical Manual (January 2000)
and the ADFM. Both setups must be the same.
F5
to view the
SETUP MENU
Up/Down
arrow keys to select the parameter to change.
Enter
Space
or
keys to change the parameter.
Escape
to exit the
SETUP MENU
F6
. The configuration file is saved to the directory where
BBTALK Communication Setup Menu
.
.
Software Operation
End
to send the
page 3-5