Chapter 6
6-2

Equipment Required

Special test equipment is not needed for troubleshooting (fault isolation). A
list of equipment required for the performance tests is listed below. Any
equipment satisfying the critical specification listed may be used.
Table 6-2:
Required Test
Equipment
Digital Multi-Meter
page 6-2
Required Test Equipment
Critical Specification
3 ½ digit
DC-Voltage Range: 200.0 mV, 2.000V, 20.00 V, 200 .0V
DC-Voltage Resolution: 100 uV, 1 mV, 10 mV, 100 mV, 1 V respective
DC-Voltage Accuracy: ± 1%
AC-Voltage Range: 200 V, 450 V
AC-Voltage Resolution: 10 mV @ 20 V range
AC-Voltage Resolution: 1 V @ 450 V range
AC-Voltage Accuracy: ± 2%
Resistance Range: 200, 2 k, 20 k, 200 k, 20 Mohm
Res.-Resolution: 0.1, 1.0, 10, 100, 1k, 10k respective
Res.-Accuracy: ± 2% @ 200 Ohm to 200 kOhm
Res.-Accuracy: ± 5% @ 20 Mohm
MGD Technologies, Inc.