Appendix C
C-2.4
Performance and Test Commands
The ADFM uses the following commands for testing.

PA - User Interactive Tests

Purpose:
Format:
Range:
Default:
Description:
Example:
>pa0
User Interactive, Built In Tests
--------------------------------
PA0 = Help
PA1 = Sensor Data
>pa1
Press any key to quit sensor display ...
Transducer Temp
21.47øC
21.47øC
21.47øC
21.47øC
21.47øC
21.47øC
21.47øC
21.47øC

PC - Diagnostic Tests

Purpose:
Format:
Description:
Example:
>pc
Transducer Communications:
Recorder BIT (RT ):
Modem (MO ):
System Voltages (PT2):
CPU RAM (PI ):
page C-22
Displays results of the ADFM sensor data diagnostic tests.
PAn
n = 0 (help), 1 = Sensor Data
PA
This diagnostic test verifies the sensor operation. We recom-
mend you run this command before a deployment.
Depth(mm)
0
0
0
0
0
0
0
0
Sends/displays results of a series of ADFM system diagnostic
tests.
PC
These diagnostic tests check the major ADFM modules and
signal paths. We recommend you run this command before a
deployment.
PASS
PASS
PASS
PASS
PASS
MGD Technologies, Inc.