Appendix C

EE - Secondary Depth Max. Probable Error

Purpose:
Format:
Range:
Default:
Description:

EO - Secondary Depth Zero Offset

Purpose:
Format:
Range:
Default:
Description:

ER - Secondary Depth Span

Purpose:
Format:
Range:
Default:
Description:
page C-14
Sets the maximum probable error of the secondary depth sen-
sor.
EEnnnn
nnnn = 0 to 1000 (o to 100%)
EE0020 (2%)
The EE command allows the user to set the maximum prob-
able error (MPE) of the secondary depth sensor. If the depth
as determined by the vertical beam is within the MPE of the
secondary sensor, the vertical beam depth is used. Otherwise,
the secondary depth sensor depth is used. The secondary
depth sensor must be selected with EZ=x1xxxxx.
Sets the secondary depth zero offset in the ADFM.
EOnnnn
8 to 9999 mm
EO0008
The secondary depth zero offset determines the distance above
the bottom of the channel at which the secondary depth sensor
is placed. The default setting of 8 mm is half the diameter of
the recommended secondary sensor.
Sets the operating range of the secondary depth sensor.
ERnnnnn
0 to 30000 mm
ER00000
Sets the full-scale depth of the secondary depth sensor in mil-
limeters. If this value is unchanged (i.e. left at zero), all
depths will be zero + zero offset (EO-command).
MGD Technologies, Inc.