ES - Salinity

Purpose:
Format:
Range:
Default:
Description:

ET - Temperature

Purpose
Format
Range
Default
Description
Example
ET = 14.00 × 100 = 1400 = ET1400 (+ is understood)
ET = -3.50 × 100 = -350 = ET-0350

EX - Coord Transform

Purpose:
Format:
Range:
Default:
Description:
ADFM Technical Manual (January 2000)
Sets the water's salinity value.
ESnn
nn = 0 to 40 parts per thousand
ES00
ES sets the water's salinity value. The ADFM uses ES in its
speed of sound calculations. The ADFM assumes the speed of
sound reading is taken at the transducer head.
Sets the water's temperature value.
ET±nnnn
±nnnn = -5.00 C to +40.00 C
ET+2500
ET sets the temperature value of the water. The ADFM uses
ET in its speed of sound calculations. The ADFM assumes
the speed of sound reading is taken at the transducer head.
Convert temperatures of +14 C and -3.5 C to ET-command
values.
Sets the coordinate transformation processing flags.
EXnnnnn
Firmware switches (see description)
EX00000
EX sets firmware switches that control the coordinate trans-
formation processing for velocity and percent-good data.