Appendix C
CK - Keep Parameters
Purpose:
Format:
Description:
CL – Leapfrog
Purpose:
Format:
Range:
Default:
Description:
CN - Site Number
Purpose:
Format:
Range:
Description:
CP - Clear BIT Log
Purpose:
Format:
Description
CR - Retrieve Parameters
Purpose:
Format:
Range:
Description:
page C-10
Stores current parameters to EEPROM.
CK
CK saves the current user command parameters to an
EEPROM on the CPU board. The ADFM maintains data
stored in the EEPROM (user settings) even if power is lost. It
does not need a battery. You can recall parameters stored in
EEPROM with the CR0-command
Prevents the ADFM from going to sleep.
CLn
n = 0 or 1 (0 = on, 1 = off)
CL1
The leapfrog command prevents the ADFM from going to
sleep. Whenever the unit tries to sleep, it will instead wait in a
non-power saving state.
Identifies the ADFM site number.
CNaaaaaaaa
8-character string
An 8-character string identifying the ADFM.
Clears the BIT Log stored in EEPROM on the MainBoard.
CP
Clears the BIT Log stored in EEPROM on the MainBoard.
The log is updated whenever an error occurs.
Resets the ADFM command set to EEPROM or factory set-
tings.
CRn
n = 0 to 1 (see description)
The ADFM automatically stores the last set of commands
used in battery-backed RAM. The ADFM will continue to be
MGD Technologies, Inc.