Vn - Vertical Beam Number Of Depth Cells; Vp - Pings Per Ensemble; Vs - Vertical Beam Depth Cell Size; Vx - Vertical Beam Transmit Length - Teledyne ADFM Pro20 Technical Manual

Acoustic doppler flow meter
Hide thumbs Also See for ADFM Pro20:
Table of Contents

Advertisement

VN - Vertical Beam Number Of Depth Cells

Purpose:
Format:
Range:
Description:

VP - Pings per Ensemble

Purpose:
Format:
Range:
Default:
Description:

VS - Vertical Beam Depth Cell Size

Purpose:
Format:
Range:
Description:

VX - Vertical Beam Transmit Length

Purpose:
Format:
Range:
Description:
ADFM Technical Manual (January 2000)
deep tracking. This allows the ADFM to change with condi-
tions during the course of the deployment.
Sets the number of depth cells over which the ADFM collects
data for the vertical beam.
VNnnn
nnn = 001 to 128 depth cells
The range of the ADFM is set by the number of depth cells
(VN) times the size of each depth cell (VS).
Sets the number of pings for the vertical beam to average in
each data ensemble.
VPnnnnn
nnnnn = 0 to 16384 pings
VP00010
VP sets the number of vertical beam pings to average in each
ensemble before sending/recording the water-track data.
Selects the volume of water for one measurement cell for the
vertical beam
VSnnnn
nnnn = 0 to 9999 cm
The ADFM collects data over a variable number of depth
cells. VS sets the size of each cell in vertical centimeters
Sets the vertical beam transmit length in centimeters.
VXnn
nn = 0 to 99 cm
Sets the vertical beam transmit length in centimeters.